7 software update – VEGA VEGAFLEX 82 Foundation Fieldbus User Manual
Page 65

65
9 Diagnostics and service
VEGAFLEX 82 • Foundation Fieldbus
44221-EN-130910
7. Cable: Fasten the cable with three pins, torque 20 Nm
(14.75 lbf in)
Cable ø 4: 7 Nm (5.16 lbf ft)
Cable ø 6, cable ø 8: 20 Nm (14.75 lbf ft)
8. Enter new probe length and then carry out a fresh adjustment
(see "Setup procedure, Carrying out min. adjustment - Carrying
out max. adjustment").
3
1
2
2
A
B
45 mm (1.77"
)
150 mm (5.91"
)
70 mm (2.76"
)
150 mm (5.91"
)
Fig. 51: Shortening the cable probe
A Gravity weight - cable ø 4 mm
B Gravity weight - cable ø 6 mm
1 Pins
2 Thread M12 for eye-bolt
3 Pins
9.7 Software update
The following components are required to update the sensor soft-
ware:
•
Sensor
•
Voltage supply
•
Interface adapter VEGACONNECT
•
PC with PACTware
•
Current sensor software as file
You can find the actual sensor software as well as detailed infor-
mation of the procedure under "
"Software".
You can find information about the installation in the download file.
Caution:
Instruments with approvals can be bound to certain software ver-
sions. Therefore make sure that the approval remains effective with a
software update.