4 Mounting

VEGAFLEX 82 • Foundation Fieldbus

44221-EN-130910

1

2

1

2

Fig. 11: Fasten the probe
1 Probe

2 Retaining sleeve

In case of difficult installation conditions in liquid applications, the

probe can be also mounted laterally. For this purpose, adapt the rod

with rod extensions or bow-shaped segments.
Let the probe length determine automatically by the instrument to

compensate the resulting running time changes.
The determine probe length can deviate from the actual probe length

when using bow-shaped segments.
If installations such as struts, ladders, etc. exist on the vessel wall,

then the probe should have a distance to the vessel wall of at least

300 mm (11.81 in).
You can find further information in the supplementary instructions of

the rod extension.

In case of difficult installation conditions, for example in a socket, the

probe can be adapted respectively with a rod extension.
Let the probe length determine automatically by the instrument to

compensate the resulting running time changes.
You can find further information in the supplementary instructions of

the rod extension.

Lateral installation

Rod extension