6 Set up with the display and adjustment module
VEGAFLEX 82 • Foundation Fieldbus
44221-EN-130910
In this menu item you can select which type of medium you want to
measure. You can choose between liquid or bulk solid.
In this menu item you can select the application. You can choose
between metallic oder non-metallic vessels.
Note:
The selection of the application has a considerable influence on all
other menu items. Keep in mind that as you continue with the param-
eter adjustment, individual menu items are only optionally available.
You have the option of choosing the demonstration mode. This mode
is only suitable for test and demonstration purposes. In this mode, the
sensor ignores the parameters of the application and reacts immedi-
ately to each change.
In this menu item, you can define the type of medium (product).
This menu item is only available if you have selected level measure-
ment under the menu item "Application".
You can choose between the following medium types:
Dielectric con-
stant
Product type
Examples
> 3
Cereals, flour
All kind of cereals, wheat flour
1.5 … 3
Granules, cement Lime, gypsum, cement
< 1.5
Dusts, wood chips Wood chips, sawdust
In this menu item, you can enter the max. adjustment for the level.
Adjust the requested percentage value with [+] and store with [OK].
Setup - Type of medium
Setup - Application
Setup - Medium, dielec-
tric constant
Setup - Max. adjustment
Level