VEGA VEGASOURCE 35 Emitter loading and exchange User Manual
Page 23

l
Insert the new source capsule (13) into the source holder (8) until it
snaps in or insert the spring ring (14) with the threaded version and
screw the source capsule into the source holder (8) (torque 2
±
0.5 Nm). Use a suitable gripping device or a manipulator for
insertion
l
Insert the source holder (8) completely into the source container
and screw in with a spanner (size 13)
l
Tighten the source holder (8)
l
Remove existing type labels (when exchanging the emitter)
l
Unpack the type label set and letter it respectively. You can find
further information in chapter "Safety instructions - Labelling"
2
1
4
3
5
2
Fig. 20: Fasten type label
1
Type label - Source holder
2
Type label - Source
3
Adhesive type label - Source
4
Adhesive label "Radioactive"
A
Arrow mark
B
Straps must line up
l
Letter the adhesive type label (48 x 30 mm) and attach it to the
source holder
l
Letter the type label (107 x 17 mm) and fasten it to the neck of the
source container with two grooved pins
l
Attach the adhesive label "Radioactive" to the specified position of
the source container (if not already present)
l
Fasten the square adapters with flange (11) with the three fixing
screws (5). The arrow mark (A) of the source holder is above the
OFF mark of the adapter disc (7)
Insert source holder
Attach or replace the
type labels
Mount squared adapters
and drive unit
Emitter loading and exchange • VEGASOURCE 31 35
23
2 Loading or emitter exchange
39155
-
EN
-110210