beautypg.com

5 instrument repair – VEGA VEGAPULS 63 (≥ 2.0.0 - ≤ 3.8) 4 … 20 mA_HART - two-wire User Manual

Page 48

background image

S

elect with the right mouse key the sensor in the project and go to

"

Additional function". Then click to "Software update". The window

"

Sensor # software update" opens. PACTware checks now the sensor

data and displays the actual hardware and software version of the
sensor. This takes approximately 60 s.

P

ush the button "Update software" and select the previously extracted

hex file. Then the software update can be started. The additional files
are installed automatically. Depending on the sensor, this procedure
lasts up to 1 h. Then the message appears ""Software update
successfully executed
".

8

.5 Instrument repair

I

f a repair is necessary, please proceed as follows:

Y

ou can download a return form (23 KB) from our Internet homepage

www.vega.com under: "Downloads - Forms and certificates - Repair
form
".

B

y doing this you help us carry out the repair quickly and without

having to call back for needed information.

l

P

rint and fill out one form per instrument

l

C

lean the instrument and pack it damage-proof

l

A

ttach the completed form and, if need be, also a safety data

sheet outside on the packaging

l

P

lease ask the agency serving you for the address of your return

shipment. You can find the respective contact data on our website
www.vega.com under: "Company - VEGA worldwide"

Load software into sen-
sor

48

VEGAPULS

63 • 4 … 20

mA/HART - two-wire

8 M

aintenance and fault rectification

28436

-EN

-120530