4 software update – VEGA VEGAPULS 63 (≥ 2.0.0 - ≤ 3.8) 4 … 20 mA_HART - two-wire User Manual
Page 47
l
A
t the factory by VEGA
l
O
r on site by the user
I
n both cases, the sensor serial number is necessary. The serial
numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.
Information:
W
hen loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Electronics
module").
T
he electronics modules are adapted to the respective sensor and
distinguish also in the signal output or power supply.
8
.4 Software update
T
he software version of VEGAPULS 63 can be determined as follows:
l
on the type label of the electronics
l
via the indicating and adjustment module
l
via PACTware
Y
ou can view all software histories on our website www.vega.com.
M
ake use of this advantage and get registered for update information
via e-mail.
T
he following components are required to update the sensor software:
l
S
ensor
l
V
oltage supply
l
VEGACONNECT
l
PC
with PACTware
l
C
urrent sensor software as file
A
t "www.vega.com/downloads" go to "Software". Select under "plics
sensors and instruments", "Firmware updates" the respective instru-
ment series and software version. Load the zip file via the right mouse
key with "Save target as" e.g. on the desktop of your PC. Move with the
right mouse key to the folder and select "Extract all". Save the
extracted files, for example on the desktop.
C
onnect the signal conditioning instrument to power supply and
provide the connection from the PC to the instrument via the interface
adapter. Start PACTware and go via the menu "Project" to the VEGA
project assistant. Select "USB" and "Set instruments online". Activate
the project assistant with "Start". The assistant establishes the
connection automatically and opens the parameter adjustment window
"
Sensor # online parameter adjustment". Connect this parameter
adjustment window before you carry out further steps.
Assignment
Load sensor software to
PC
Prepare update
VEGAPULS
63 • 4 … 20
mA/HART - two-wire
47
8 M
aintenance and fault rectification
28436
-EN
-120530