beautypg.com

VEGA VEGABAR 61 Foundation Fieldbus User Manual

Page 58

background image

-

Flange 1" 150 lbs RF ANSI B16.5; with
extension 2", 3"

1.8

-

Flange 2" 150 lbs RF ANSI B16.5; with
extension 2"

2

Temperature coefficient of a cooling ele-
ment, depending on the diaphragm-ø

0.1 … 1.5

Temperature coefficient of a 1 m long
capillary line, depending on the diaphragm-ø

0.1 … 15

Process conditions

The specifications to the pressure stage and the product temperature are used as an overview.
The specifications of the type label are applicable.

Product temperature depending on the isolating liquid (temperature: p

abs

>

1 bar/14.5 psi/p

abs

<

1 bar/14.5 psi)

14)

-

silicone oil KN2.2

-40 … +150 °C/-40 … +150 °C
(-40 … +302 °F/-40 … +302 °F)

-

Silicone oil KN2.2 and cooling element
or capillaries

-40 … +200 °C/-40 … +150 °C
(-40 … +392 °F/-40 … +302 °F)

-

Vegetable oil

-10 … +130 °C/-10 … +130 °C (14 … +266 °F/
14 … +266 °F)

-

Water and glycerine

-10 … +120 °C (+14 … +248 °F)

-

High temperature oil KN3.2 and cooling
element

-10 … +200 °C/-10 … +200 °C (+14 … +392 °F/
+

14 … +392 °F)

-

High temperature oil KN3.2 and cooling
element 300 mm or capillaries

-10 … +400 °C/-10 … +200 °C (+14 … +752 °F/
+

14 … +392 °F)

-

Halocarbon oil KN21

-40 … +150 °C/-40 … +80 °C
(-40 … +302 °F/-40 … +176 °F)

-

Halocarbon oil KN21 for oxygen appli-
cations

-40 … +60 °C/-40 … +60 °C
(-40 … +140 °F/-40 … +140 °F)

-

Silicone-free liquid KN70

-40 … +70 °C (-40 … +158 °F), no vacuum

-

med. white oil KN62 and cooling element

-15 … +150 °C/-15 … +150 °C (+5 … +302 °F/
+

10 … +302 °F)

-

Med. white oil KN62 (FDA) and cooling
element

-15 … +200 °C/-15 … +150 °C (+5 … +392 °F/
+

10 … +302 °F)

-

Med. white oil KN62 (FDA) and temper-
ature adapter 300 mm

-15 … +250 °C/-15 … +170 °C (+5 … +482 °F/
+

10 … +338 °F)

-

Med. white oil KN62 (FDA) and capil-
laries 3 m

-15 … +250 °C/-15 … +170 °C (+5 … +482 °F/
+

10 … +338 °F)

Vibration resistance

mechanical vibrations with 4 g and 5 … 100 Hz

15)

Shock resistance

Acceleration 100 g/6 ms

16)

14)

Version for oxygen applications up to 60 °C (140 °F).

15)

Tested according to the regulations of German Lloyd, GL directive 2.

16)

Tested according to EN 60068-2-27.

58

VEGABAR

61 • Foundation Fieldbus

10 Supplement

28258

-EN

-081209