1.7 Technical data

Power supply

Supply voltage

9 … 32 V DC)

(output voltage U

O

of the segment

dependent on the segment coupler used

coupler)

see PA specification, e.g.

- non Ex

e.g. 22 V DC (nominal voltage of the segment

coupler) max. 32 sensors on one two-wire
cable

- E x

13.5 V DC nominal voltage of the segment
coupler max. 10 sensors on one two-wire
cable (typical 8 sensors)

Current consumption

constant 10 mA (no leakage current output)

Cable load

dependent on the segment coupler, see
technical data of the segment couplers and
Profibus specification

Measuring range

VEGAFLEX 51

0.3 … 10 m

VEGAFLEX 52

0.3 … 20 m

Output signal

digital (Profibus)

two-wire technology:

the digital output signal (meas. signal is
modulated onto the power supply and is further
processed in the PLC or in the process control

Integration time

0 … 999 seconds (adjustable)

Adjustment

- adjustment software VEGA Visual Operating on Master-Class 2 PC
- adjustment module MINICOM in the sensor or in the external indicating instrument

(optional)

- process adjustment interface PACTware

TM

(VVO as subprogram)

- SIMATIC PDM in conjunction with Electronic Device Description (EDD)

Accuracy

1)

(under reference conditions acc. to IEC 770 - relating to the max. measuring range)

Linearity error

< 0.1 %

Temperature drift

0.015 %/10 K

Characteristics

linear

Deviation in characteristics including

linearity, reproducibility and
hysteresis (determined acc. to the
limit point method)

better than 0.05 %

Average temperature coefficient of the
zero signal

0.06 %/10 K

Resolution general

max. 1 mm

Resolution of the output signal

0.005 % or 1 mm

Technical data