Vessel without cathodic corrosion protection, Electrical connection – VEGA Capacitive electrodes EL … User Manual

Page 32

background image

32

Capacitive electrodes EL …

Capacitive electrode with external overvoltage protection unit

Vessel without cathodic corrosion protection

E2

E1

A2

A1

A1

A2

E2

E1

1
2

Typ

B62-36G

Typ

B62-36G

B

A

Ex-area

Not-Ex-area (control
room)

either overvoltage arrester

either overvoltage arrester
B62 - 36 G Ex 0 (metal housing) or
B62 - 36 G Ex (plastic housing)
(L

i

= 0,15 mH, C

i

= 1,5 nF)

Electrode EL

Outer
isolation

Outer
isolation

Screen

Screen

Signal conditioning
instrument or
safety barrier

Earth / PA-terminal

to potential
equalisation line

External earth terminal

Note:
[A] Between control room and overvoltage protection system a suitable cable, if required with screen or metal

cover, should be used. Screen or metal cover - if required - must be only connected to the overvoltage
protection system on the electrode side.

[B] Between overvoltage protection system and capacitive electrode a suitable cable with metal cover, screen or

a suitable cable with metal protection tube must be used (screen, metal cover or protective tube must be
connected to the potential equalization).

Test voltage of cable A and B:

≥ 500 V AC

min. 4 mm

2

Cu

Vessel
Zone 0

Electrical connection