VEGA VEGABAR 17 User Manual
Page 27
A
verage temperature coefficient of the span
typ. < 0.2 %/10 K
Long-term stability (according to DIN 16086, DINV 19259-1 and IEC 60770-1)
L
ong-term drift of the zero signal
6)
<
0.2 %/
year
Ambient conditions
A
mbient temperature (note temperature derating!)
-
C
able outlet
-20 … +80 °C (-4 … +176 °F)
-
R
ound plug connector M12 x 1
-25 … +80 °C (-13 … +176 °F)
-
A
ngled plug connector according to
ISO
4400
-40 … +80 °C (-40 … +176 °F)
-
T
erminal housing
-50 … +80 °C (-58 … +176 °F)
-
with cooling element
-20 … +80 °C (-4 … +176 °F)
2
1
55°C
(131°F)
80°C
(176°F)
130°C
(266°F)
150°C
(302°F)
Fig. 14: Temperature derating VEGABAR 17
1
Ambient temperature
2
Process temperature
S
torage and transport temperature
-
S
tandard
-30 … +100 °C (-22 … +212 °F)
-
with cooling element
-20 … +100 °C (-4 … +212 °F)
Process conditions
P
roduct temperature
-
S
tandard
-30 … +100 °C (-22 … +212 °F)
-
additional
-30 … +125 °C (-22 … +257 °F)
-
with cooling element
-20 … +150 °C (-4 … +302 °F)
-
M
easuring ranges from 400 bar
-30 … +70 °C (-22 … +158 °F)
S
hock resistance
-
V
ersion with terminal housing
600
g according to IEC 60068-2-27 (mechanical
shock)
6)
U
nder reference conditions, relating to the adjusted span.
VEGABAR
17
27
9 S
upplement
27636
-
EN
-120531