6 setup, 1 setup steps, 2 recalibration – VEGA VEGABAR 17 User Manual
Page 18: 6setup
6
Setup
6
.1 Setup steps
A
fter mounting and electrical connection, VEGABAR 17 is ready for
operation.
VEGABAR
17
delivers a current of 4 … 20 mA corresponding to the
actual process pressure.
F
urther settings are not necessary.
6
.2 Recalibration
W
ith both instruments with thread ring or field housing, zero and span
can be adjusted via integrated potentiometers. Adjustment range:
l
Z
ero ± 5 %
l
S
pan ± 5 %
T
his allows, for example, the consideration of an installation position
different from the reference installation position.
A
shifting of zero shifts span also respectively.
Note:
T
he potentiometer for span should only be used if you have adequate
calibration equipment (at least 3 times more precise than the deviation
of VEGABAR 17).
R
ecommended recalibration cycle: 1 year.
P
roceed as follows:
1
L
oosen the plug connector and screw the screwed ring in
connected status
2
P
lace the plug connector onto the instrument place and pull both
carefully out of the instrument
Instruments with plug
connector or cable out-
let
18
VEGABAR
17
6 S
etup
27636
-EN
-120531