3 auto-store setup – Soft dB Opus Suite SLM 4-Ch Module User Manual
Page 42
39
8.3 Auto-Store Setup
The Auto-Store Setup provides several options for saving measurements.
AutoStore Modes
The measurement will stop after the Leq Average Time elapses. The
resulting DataSet must be stored manually by pressing the SAVE button.
The measurement will stop after the Leq Average Time elapses. The
resulting DataSet is stored automatically.
The measurement continues after the Leq Average Time elapses, averaging
is reset and each period data is stored automatically.
Multiple Mode Setup
The Align Leq Period on Clock Time forces the alignment of the Leq periods
to match the time on the clock.
This control is only visible with ID increment set to On Start Only or Every
‘X’ Leq. The alignment with clock is always forced when Every
Hour/Day/Week modes are selected.
St
op Mo
de
The measurement will stop when it is stopped manually by the user.
The measurement will stop automatically when the fixed duration has
elapsed. It can also be stopped manually by clicking the STOP button.
The measurement will stop automatically when the stop time is reached. It
can also be stopped by clicking the STOP button.