Ohaus RA PRICE COMPUTING COMPACT SCALES Manual User Manual
Page 14

EN-12
RA Series
6. When “
YES
” is displayed, press [+] to enter the Span Calibration
sequence.
Note: Do not change the settings of any parameters unless specified
above. Other parameters are service settings intended for authorized 
servicing use only. The scale may be turned off at any time to abort setup 
without saving any changes. 
 
7. Span Calibration Sequence: 
 
Window Display
Weight
Unit Price
Total Price
Description
Actions
CAL StEP
1 15.00
Scale capacity
Press [+] to select and proceed to next step. 
Make sure the displayed value matches the 
scale model capacity. 
Do not press [>T<], which will toggle 
through other capacity settings that may not 
match the scale model’s.
Step
2 ------
Establish zero
Press [+] to initiate countdown sequence. 
Ensure the pan is empty and the scale 
stable.
5 ... 0
Countdown
The countdown will repeat if the scale is not 
stable during this process. 
Step
3 kkkkkk
2/3 calibration 
point 
Place 2/3 capacity weight on the pan, then 
press [+] to proceed to the next step.
Step
4 0
Full span
Enter the scale capacity value from Step 1 
(e.g., “
15
”) via the numeric keypad.
15
Establish full 
span 
Place the indicated weight on the pan, 
then press [+] to initiate countdown 
sequence.
5 ... 0
Countdown
The countdown will repeat if the scale is not 
stable during this process. 
Save
Save or Abort
Remove the weight from the pan, then 
press [+] to save settings and restart scale 
to weighing mode. 
If needing to abort calibration and retain 
previous calibration data, press [>T<] 
(“Abort” will be displayed) then [+] at this 
time (or turn the scale off at any time during 
the process).