Monitoring Accuracy*
Pressure
readings:
±2 cm H
2
0
Volume
readings:
±10% or ±15 mL, whichever is greater
O
2
concentration
monitor:
±3%
Delivery Accuracy*
Inspired
pressure
control:
±2 cm H
2
O
Oxygen – Air
mixing:
±3% V/V of setting
Tidal volume
delivery:
±10% of setting or ±5 mL,
whichever is greater
Nebulization
Nebulizer:
Aeroneb Nebulizer System built-in
Nebulizer
technology:
Electronic micro pump
Nebulizer
run time:
10, 15, 20 or 30 minutes
Auto-repeat
capability:
Cycles: 1 to 10
Pause Time: 30 sec to 8 hr
1 to 5 minutes (increments of 1 minute)
5 to 55 minutes (increments of 5 minutes)
1 to 8 hours (increments of 0.5 hour)
Nebulizer
volume
setting:
2.5, 3, 5 or 6 mL
Particle size:
Aeroneb Pro: average 2.1 microns MMAD
(Mean Mass Aerodynamic Diameter)
Aeroneb Solo: 3.4 microns MMAD
Residual
volume:
Aeroneb Pro: average 0.3 mL
Aeroneb Solo: average < 0.1 mL
Performance may vary depending upon the type of drug used.
For additional information contact Aerogen or drug supplier.
Pnuematic nebulizer
Flow
compensation: 1 to 12 L/min (increments of 0.5 L/min)
Trends
Trend data:
All set parameters and measured data
Trend styles:
Measured and graphic
Maximum
trending:
14 days (336 hours)
Trend
scaling:
12 min, 1h, 2h, 4h, 6h, 8h, 10h, 12h, 24h, 36h,
48h and 72h
Resolution:
1 minute intervals for most recent 12 hours,
5 minute intervals for 12 to 48 hours,
30 minute intervals after 48 hours
Mini-Trends:
Waveform values can be displayed as a trend
in a split screen view
Mini-Trends
parameters
are based on
the waveform
displayed:
Paw (P
peak
, P
plat
or Leak)
Flow
(MV
exp
, RR)
Volume (Spont MV or Mech MV,
Spont RR or Mech RR)
P
aux
(P
peak
)
External Communications
Communication
ports:
Serial port (RS-232), RS-485 port, RS-422 port,
1 USB port, Ethernet port, Compact fl ash
card socket, nurse call
EView (optional)
Data Available: 10 snapshots
7 days of vent data
Optional breath to breath waveform data
Electrical Specifi cations
Line supply
Line voltage:
85 to 132 Vac, 47/63 Hz
190 to 264 Vac, 47/63 Hz
Power
consumption:
< 200 W
* Ventilation delivery specifi cations requirements:
• Operating at EN794 and ASTM F1100 patient conditions
• Operating at 21°C and at 1000 mbar ambient pressure
• All volumes are at ambient temperature and pressure, dry (ATPD)