Monitoring Accuracy*

Pressure
readings:

±2 cm H

2

0

Volume
readings:

±10% or ±15 mL, whichever is greater

O

2

concentration
monitor:

±3%

Delivery Accuracy*

Inspired
pressure
control:

±2 cm H

2

O

Oxygen – Air
mixing:

±3% V/V of setting

Tidal volume
delivery:

±10% of setting or ±5 mL,

whichever is greater

Nebulization

Nebulizer:

Aeroneb Nebulizer System built-in

Nebulizer
technology:

Electronic micro pump

Nebulizer
run time:

10, 15, 20 or 30 minutes

Auto-repeat
capability:

Cycles: 1 to 10

Pause Time: 30 sec to 8 hr

1 to 5 minutes (increments of 1 minute)

5 to 55 minutes (increments of 5 minutes)

1 to 8 hours (increments of 0.5 hour)

Nebulizer
volume
setting:

2.5, 3, 5 or 6 mL

Particle size:

Aeroneb Pro: average 2.1 microns MMAD

(Mean Mass Aerodynamic Diameter)

Aeroneb Solo: 3.4 microns MMAD

Residual
volume:

Aeroneb Pro: average 0.3 mL

Aeroneb Solo: average < 0.1 mL

Performance may vary depending upon the type of drug used.
For additional information contact Aerogen or drug supplier.

Pnuematic nebulizer

Flow
compensation: 1 to 12 L/min (increments of 0.5 L/min)

Trends

Trend data:

All set parameters and measured data

Trend styles:

Measured and graphic

Maximum
trending:

14 days (336 hours)

Trend
scaling:

12 min, 1h, 2h, 4h, 6h, 8h, 10h, 12h, 24h, 36h,

48h and 72h

Resolution:

1 minute intervals for most recent 12 hours,

5 minute intervals for 12 to 48 hours,

30 minute intervals after 48 hours

Mini-Trends:

Waveform values can be displayed as a trend

in a split screen view

Mini-Trends
parameters
are based on
the waveform
displayed:

Paw (P

peak

, P

plat

or Leak)

Flow

(MV

exp

, RR)

Volume (Spont MV or Mech MV,

Spont RR or Mech RR)

P

aux

(P

peak

)

External Communications

Communication
ports:

Serial port (RS-232), RS-485 port, RS-422 port,

1 USB port, Ethernet port, Compact fl ash

card socket, nurse call

EView (optional)

Data Available: 10 snapshots

7 days of vent data

Optional breath to breath waveform data

Electrical Specifi cations

Line supply

Line voltage:

85 to 132 Vac, 47/63 Hz

190 to 264 Vac, 47/63 Hz

Power
consumption:

< 200 W

* Ventilation delivery specifi cations requirements:

• Operating at EN794 and ASTM F1100 patient conditions

• Operating at 21°C and at 1000 mbar ambient pressure

• All volumes are at ambient temperature and pressure, dry (ATPD)

EngstromProROWSpecSheet_DOC0969480_eng.indd 7