Ventilator Preferences

Back-up
Mode:

Establishes the specifi c ventilator mode
and parameters used in the event that the
ventilator switches to Back-up ventilation

ARC:

Allows control and setting of the airway
resistance compensation

Assist
Control:

Allows the user to turn the Assist Control
capability On or Off

Leak
Compensation: Allows the user to turn the Leak

Compensation capability On or Off

Trigger
Compensation: Allows the user to turn On or Off

compensation for fl ow triggering

TV Based
Conditions:

Allows setting between ATPD (Ambient
Temperature Pressure Dry) or BTPS
(Body Temperature Pressure Saturated)

Airway Resistance Compensation (ARC)

Type of
compensation: Electronic tube compensation

Compensation
for:

Endotracheal and tracheostomy tubes

Tube
diameter:

5 to 10 mm

Level of
compensation: 25 to 100%

Mode Families

Allows user adjustment to specify certain parameters that
align with the hospital’s current ventilator usage.

Adjustable
parameters:

Flow and Inspiratory timing

Family 1:

Flow control is On/Insp. Timing is I:E

Family 2:

Flow control is Off/Insp. Timing is I:E

Family 3:

Flow control is On/Insp. Timing is T

insp

Family 4:

Flow control is Off/Insp. Timing is T

insp

Family 5:

Flow control is On/Insp. Timing is T

pause

Help

Help screens are available for any recent or active alarm.
Help will provide the cause of the alarm and the action to
resolve the alarm conditions.

Ventilator Monitoring

Airway
pressure:

–20 to +120 cm H

2

O

Patient fl ow:

1 to 200 L/min

Tidal volume:

5 to 2,500 mL

Minute volume: 0 to 99.9 L/min

Compliance:

0.1 to 150 mL/cm H

2

O

Resistance:

1 to 500 cm H

2

O/L/s

Rate:

0 to 120 breaths per minute

(increments of 1 breath per minute)

FiO

2

:

10 to 100%

Rapid Shallow
Breathing
Index (RSBI):

1 to 999 bpm/L

Oxygen Monitoring

Technology:

Dynamic Paramagnetic Oxygen

monitoring

system

Life span:

Unlimited operating life due to the use of

non-depleting

technology

Screen

Display type:

30.5 cm/12 inch touch screen full color LCD
adjustable viewing angle

Waveforms
in screen:

Three at a time

Waveform
parameters:

Pressure, fl ow, volume, CO

2

, O

2

and auxiliary pressure

Graphic
scaling:

Automatic scaling for optimal size or

independent

scaling

Data:

Control parameters, patient data, alarm
settings and messages

Status
indicator:

Ventilation mode, battery level, clock

Favorites:

23 Hyperlink shortcuts to choose from

7 selectable at one time

EngstromProROWSpecSheet_DOC0969480_eng.indd 6

EngstromProROWSpecSheet_DOC0969480_eng.indd 6

26.5.2011 14.10