Ventilator preferences, Airway resistance compensation (arc), Mode families – GE Healthcare Engström Pro User Manual
Page 6: Help, Ventilator monitoring, Oxygen monitoring, Screen
Ventilator Preferences
Back-up
Mode:
Establishes the specifi c ventilator mode
and parameters used in the event that the
ventilator switches to Back-up ventilation
ARC:
Allows control and setting of the airway
resistance compensation
Assist
Control:
Allows the user to turn the Assist Control
capability On or Off
Leak
Compensation: Allows the user to turn the Leak
Compensation capability On or Off
Trigger
Compensation: Allows the user to turn On or Off
compensation for fl ow triggering
TV Based
Conditions:
Allows setting between ATPD (Ambient
Temperature Pressure Dry) or BTPS
(Body Temperature Pressure Saturated)
Airway Resistance Compensation (ARC)
Type of
compensation: Electronic tube compensation
Compensation
for:
Endotracheal and tracheostomy tubes
Tube
diameter:
5 to 10 mm
Level of
compensation: 25 to 100%
Mode Families
Allows user adjustment to specify certain parameters that
align with the hospital’s current ventilator usage.
Adjustable
parameters:
Flow and Inspiratory timing
Family 1:
Flow control is On/Insp. Timing is I:E
Family 2:
Flow control is Off/Insp. Timing is I:E
Family 3:
Flow control is On/Insp. Timing is T
insp
Family 4:
Flow control is Off/Insp. Timing is T
insp
Family 5:
Flow control is On/Insp. Timing is T
pause
Help
Help screens are available for any recent or active alarm.
Help will provide the cause of the alarm and the action to
resolve the alarm conditions.
Ventilator Monitoring
Airway
pressure:
–20 to +120 cm H
2
O
Patient fl ow:
1 to 200 L/min
Tidal volume:
5 to 2,500 mL
Minute volume: 0 to 99.9 L/min
Compliance:
0.1 to 150 mL/cm H
2
O
Resistance:
1 to 500 cm H
2
O/L/s
Rate:
0 to 120 breaths per minute
(increments of 1 breath per minute)
FiO
2
:
10 to 100%
Rapid Shallow
Breathing
Index (RSBI):
1 to 999 bpm/L
Oxygen Monitoring
Technology:
Dynamic Paramagnetic Oxygen
monitoring
system
Life span:
Unlimited operating life due to the use of
non-depleting
technology
Screen
Display type:
30.5 cm/12 inch touch screen full color LCD
adjustable viewing angle
Waveforms
in screen:
Three at a time
Waveform
parameters:
Pressure, fl ow, volume, CO
2
, O
2
and auxiliary pressure
Graphic
scaling:
Automatic scaling for optimal size or
independent
scaling
Data:
Control parameters, patient data, alarm
settings and messages
Status
indicator:
Ventilation mode, battery level, clock
Favorites:
23 Hyperlink shortcuts to choose from
7 selectable at one time
EngstromProROWSpecSheet_DOC0969480_eng.indd 6
EngstromProROWSpecSheet_DOC0969480_eng.indd 6
26.5.2011 14.10
26.5.2011 14.10