Control and Ranges
(continued)
Incremental
settings:
20 to 50 mL (increments of .5 mL)
50 to 100 mL (increments of 1 mL)
100 to 300 mL (increments of 5 mL)
300 to 1000 mL (increments of 25 mL)
1000 to 2000 mL (increments of 50 mL)
For VCV, PCV-VG, SIMV-VC, SIMV-PCVG
and
BiLevel-VG
Patient weight: 5 to 15 kg (increments of 0.5 kg)
15 to 100 kg (increments of 1 kg)
100 to 200 kg (increments of 2 kg)
10 to 34 lb (increments of 1 lb)
34 to 220 lb (increments of 1 lb)
220 to 440 lb (increments of 5 lb)
Inspiratory
pressure
(P
insp
) range:
1 to 98 cm H
2
O (increments of 1 cm H
2
O)
P
high
:
1 to 98 cm H
2
O (increments of 1 cm H
2
O)
P
low
:
Off, 1 to 50 cm H
2
O (increments of 1 cm H
2
O)
Pressure limit
(P
limit
) range:
7 to 100 cm H
2
O for VCV and SIMV-VC
(increments of 1 cm H
2
O)
Max. inspiratory
pressure
(P
max
) limit:
7 to 100 cm H
2
O (increments of 1 cm H
2
O)
9-100 cm H
2
O (increments of 1 cm H
2
O) in NIV
PEEP:
Off, 1 to 50 cm H
2
O (increments of 1 cm H
2
O)
2-20 cm H
2
O (increments of 1 cm H
2
O) in NIV
Inspiratory
time:
0.25 to 15 sec
0.25 to 1 sec (increments of 0.05)
1 to 4 sec (increments of 0.10)
4 to 15 sec (increments of 0.25)
T
high
:
0.25 to 15 sec
0.25 to 1 sec (increments of 0.05)
1 to 4 sec (increments of 0.1)
4 to 15 sec (increments of 0.25)
T
low
:
0.25 to 18 sec
0.25 to 1 sec (increments of 0.05)
1 to 4 sec (increments of 0.1)
4 to 18 sec (increments of 0.25)
T
supp
:
0.25 to 4 sec for NIV
0.25 to 1 sec (increments of 0.05)
1 to 4 sec (increments of 0.1)
Expiratory
time:
0.25 to 59.75 sec
0.5 to 59.75 sec in NIV
Rise time:
0 to 500 ms of inspiratory period for either
fl ow or pressure depending on the mode
selected. Active in VCV, PCV, PCV-VG, SIMV-VC,
SIMV-PC, SIMV-PCVG, BiLevel-VG, NIV and
BiLevel (increments of 50 ms)
PSV rise time:
0 to 500 ms of inspiratory period for pressure
supported breaths only. Active in SIMV-VC,
SIMV-PC, SIMV-PCVG, BiLevel, BiLevel-VG and
CPAP/PSV (increments of 50 ms)
Trigger window: 0 to 80% of expiration time (increments of 5%)
Flow trigger:
1 to 9 L/min (increments of 0.5 L/min)
1 to 3 L/min (increments of 0.1 L/min)
3 to 9 L/min (increments of 0.5 L/min)
Pressure
trigger:
–10 to –3 cm H
2
O (increments of 0.5 cm H
2
O)
–3 to –0.25 cm H
2
O (increments of 0.25 cm H
2
O)
Bias fl ow
rate:
2 to 10 L/min (increments of 0.5 L/min)
8 to 20 L/min for NIV (increments of 0.5 L/min)
Insp. pause:
0 to 75% of inspiration time (increments of 5%)
T
pause
:
0 to 11 sec
0 to 1 sec (increments of 0.05)
1 to 4 sec (increments of 0.1)
4 to 11 (increments of 0.25)
Pressure
support from
PEEP level:
0 to 60 cm H
2
O for SIMV-VC, SIMV-PC,
SIMV-PCVG, BiLevel, BiLevel-VG and CPAP/PSV
(increments of 1 cm H
2
O)
0 to 30 cm H
2
O for NIV
(increments of 1 cm H
2
O)
End fl ow
level:
5 to 80% of peak fl ow for NIV, SIMV-VC, SIMV-PC,
SIMV-PCVG, BiLevel, BiLevel-VG and CPAP/PSV
(increments of 5%)
Alarm Settings
Tidal volume:
Low:
Off, 5 to 1950 mL
High:
10 to 2000 mL, Off
Minute
volume:
Low:
0.01 to 40 L/min
High:
0.4 to 99 L/min
Respiratory
rate:
Low:
Off, 1 to 99/min
High:
2 to 120/min, Off
Inspired
oxygen (FiO
2
): Low:
18 to 99%
High:
24 to 100%, Off
P
max
:
High:
7 to 100 cm H
2
O
9-100 cm H
2
O
(increments of 1 cm H
2
O) in NIV
EngstromProROWSpecSheet_DOC0969480_eng.indd 4