Control and Ranges

(continued)

Incremental
settings:

20 to 50 mL (increments of .5 mL)

50 to 100 mL (increments of 1 mL)

100 to 300 mL (increments of 5 mL)

300 to 1000 mL (increments of 25 mL)

1000 to 2000 mL (increments of 50 mL)

For VCV, PCV-VG, SIMV-VC, SIMV-PCVG

and

BiLevel-VG

Patient weight: 5 to 15 kg (increments of 0.5 kg)

15 to 100 kg (increments of 1 kg)

100 to 200 kg (increments of 2 kg)

10 to 34 lb (increments of 1 lb)

34 to 220 lb (increments of 1 lb)

220 to 440 lb (increments of 5 lb)

Inspiratory
pressure
(P

insp

) range:

1 to 98 cm H

2

O (increments of 1 cm H

2

O)

P

high

:

1 to 98 cm H

2

O (increments of 1 cm H

2

O)

P

low

:

Off, 1 to 50 cm H

2

O (increments of 1 cm H

2

O)

Pressure limit
(P

limit

) range:

7 to 100 cm H

2

O for VCV and SIMV-VC

(increments of 1 cm H

2

O)

Max. inspiratory
pressure

(P

max

) limit:

7 to 100 cm H

2

O (increments of 1 cm H

2

O)

9-100 cm H

2

O (increments of 1 cm H

2

O) in NIV

PEEP:

Off, 1 to 50 cm H

2

O (increments of 1 cm H

2

O)

2-20 cm H

2

O (increments of 1 cm H

2

O) in NIV

Inspiratory
time:

0.25 to 15 sec

0.25 to 1 sec (increments of 0.05)

1 to 4 sec (increments of 0.10)

4 to 15 sec (increments of 0.25)

T

high

:

0.25 to 15 sec

0.25 to 1 sec (increments of 0.05)

1 to 4 sec (increments of 0.1)

4 to 15 sec (increments of 0.25)

T

low

:

0.25 to 18 sec

0.25 to 1 sec (increments of 0.05)

1 to 4 sec (increments of 0.1)

4 to 18 sec (increments of 0.25)

T

supp

:

0.25 to 4 sec for NIV

0.25 to 1 sec (increments of 0.05)

1 to 4 sec (increments of 0.1)

Expiratory
time:

0.25 to 59.75 sec

0.5 to 59.75 sec in NIV

Rise time:

0 to 500 ms of inspiratory period for either
fl ow or pressure depending on the mode

selected. Active in VCV, PCV, PCV-VG, SIMV-VC,
SIMV-PC, SIMV-PCVG, BiLevel-VG, NIV and
BiLevel (increments of 50 ms)

PSV rise time:

0 to 500 ms of inspiratory period for pressure
supported breaths only. Active in SIMV-VC,
SIMV-PC, SIMV-PCVG, BiLevel, BiLevel-VG and
CPAP/PSV (increments of 50 ms)

Trigger window: 0 to 80% of expiration time (increments of 5%)

Flow trigger:

1 to 9 L/min (increments of 0.5 L/min)

1 to 3 L/min (increments of 0.1 L/min)

3 to 9 L/min (increments of 0.5 L/min)

Pressure
trigger:

–10 to –3 cm H

2

O (increments of 0.5 cm H

2

O)

–3 to –0.25 cm H

2

O (increments of 0.25 cm H

2

O)

Bias fl ow
rate:

2 to 10 L/min (increments of 0.5 L/min)

8 to 20 L/min for NIV (increments of 0.5 L/min)

Insp. pause:

0 to 75% of inspiration time (increments of 5%)

T

pause

:

0 to 11 sec

0 to 1 sec (increments of 0.05)

1 to 4 sec (increments of 0.1)

4 to 11 (increments of 0.25)

Pressure
support from
PEEP level:

0 to 60 cm H

2

O for SIMV-VC, SIMV-PC,

SIMV-PCVG, BiLevel, BiLevel-VG and CPAP/PSV
(increments of 1 cm H

2

O)

0 to 30 cm H

2

O for NIV

(increments of 1 cm H

2

O)

End fl ow
level:

5 to 80% of peak fl ow for NIV, SIMV-VC, SIMV-PC,
SIMV-PCVG, BiLevel, BiLevel-VG and CPAP/PSV
(increments of 5%)

Alarm Settings

Tidal volume:

Low:

Off, 5 to 1950 mL

High:

10 to 2000 mL, Off

Minute
volume:

Low:

0.01 to 40 L/min

High:

0.4 to 99 L/min

Respiratory
rate:

Low:

Off, 1 to 99/min

High:

2 to 120/min, Off

Inspired
oxygen (FiO

2

): Low:

18 to 99%

High:

24 to 100%, Off

P

max

:

High:

7 to 100 cm H

2

O

9-100 cm H

2

O

(increments of 1 cm H

2

O) in NIV