beautypg.com

GE Healthcare Aestiva_5 MRI anesthesia machine User Manual

Page 7

background image

Breathing circuit specifications

Operational modes

Breathing circuit
modules:

Interchangeable circle or bain (Mapleson D)

Carbon dioxide absorbent canisters (2)

Absorbent
capacity:

1.35 kg/3 lb each

Canister
release:

Integrated sensing mechanism,
CO

2

bypass capability (optional)

Ports and connectors

Exhalation:

22 mm OD ISO 15 mm ID taper

Inhalation:

22 mm OD ISO 15 mm ID taper

Bag port:

22 mm OD

Pressure gauge

Scale range:

0 to 10 kPa/–20 to 100 cm H

2

0

Bag-to-Ventilator switch

Type:

Bi-stable

Control:

Controls ventilator and direction of
breathing gas within the circuit.

Materials

All materials in contact with exhaled patient gases are
autoclavable, except standard flow sensors.

All materials in contact with patient breathing gases
are free of natural rubber latex.

Breathing circuit parameters

Compliance:

Bag mode

Mechanical mode

5.15 mL/cm H

2

O Automatically compen-

sates for compression
losses within the absorber
and bellows assembly.

Circuit volume:

5.5 L

Expiratory
resistance:

P

insp

P

exp

Flow rate

Pressure drop

Pressure drop

10 L/min

0.74 cm H

2

O

1.00 cm H

2

O

30 L/min

2.32 cm H

2

O

2.36 cm H

2

O

60 L/min

5.93 cm H

2

O

5.26 cm H

2

O

Integrated Adjustable Pressure Limiting (APL) valve

Range:

0.8 to 70 cm H

2

0

Tactile knob
indication at:

30 cm H

2

0 and above

Adjustment
range
of rotation:

0.8 to 30 cm H

2

O/0 to 230°

30 to 70 cm H

2

O/230 to 330°

Anesthetic gas scavenging

Type

Market

Hospital

Machine

system

connection

required

Active low flow: US and others

High vacuum

DISS evac

36 L/min
(300 mmHg)
@ 12 in Hg

Active low flow Japan

Adjustable

12.7 mm/

without flow

Venturi with

0.5 in

indicator:

flowmeter

hose barb

> 30 L/min

Active high flow: UK/related

Low vacuum

30 mm/1.2 in

40 - 130 L/min

BSI Male
threaded

Passive:

Germany

Venturi

25mm/0.98 in

50 L/min

hose barb

Passive:

Generic

Passive or

30 mm/1.2 in

externally

M ISO taper

attached
active system

Passive:

Sweden

Venturi/Ejector 12 mm/0.47 in

Norway

> 30 L/min

hose barb

Passive:

Denmark

Venturi/Ejector 8 mm/0.31 in
> 30 L/min

hose barb

AN3175-E.qxd 19.6.2006 12:37 Sivu 7