Electrical specifications
Current leakage
120 V:
< 300µA
220 V:
< 500µA
Power and battery back-up
Power input: USA/Canada/Mexico:
120 Vac, 60 Hz, 10A
Euro:
220-240 Vac, 50 Hz, 6A
Italy/France/Belgium:
230 Vac, 50 Hz, 6A
Japan:
100 Vac, 50 or 60 Hz, 10A
UK:
240 Vac, 50 Hz, 6A
Backup
power:
Demonstrated battery backup time under
typical operating conditions is 45 minutes
when fully charged
Battery type: Internal rechargeable sealed lead acid
Power cord: Length: 5 m/16.4 ft
Rating: 10A @ 250 Vac or 15A @ 120 Vac
Communication port
Serial
interface:
Isolated RS-232C compatible port
Inlet/outlet modules
220-240 V
100-120 V
System
circuit
breakers:
No outlets
No outlets
Pneumatic specifications
Internal common gas outlet
Connector:
ISO 22 mm OD and 15 mm ID
Gas supply
Pipeline input
range:
240 kPa to 600 kPa/35 psig to 88 psig
Pipeline
connections: DISS-male, DISS-female, DIN 13252, AS4059,
F90-116, or NIST (ISO 5359). All fittings available
for O
2
, N
2
O and Air, and contain pipeline filter
and check valve.
Cylinder
input:
Pin indexed in accordance with CGA-V-1 or
DIN (nut and gland); contains input filter and
check valve
Note: Maximum 2 cylinders of each gas;
4 cylinders total
Primary regulator
diaphragm
minimum burst
pressure:
2758 kPa/400 psig
Primary regulator
nominal output: ≤ 338 kPa/49 psig
Pin indexed cylinder connections
< 407 kPa/59 psig, DIN cylinder connections
Gas power outlet (optional)
Connector:
DISS indexed in accordance with
CGA-V-5 or Anatrir
Gas:
Oxygen
Pressure and flow
characteristics: Varies with source