Ventilator operating specifications

Ventilation operating modes

Volume Control

Pressure Control

Synchronized Intermittent Mandatory Ventilation (SIMV)
(optional)

Pressure Support (PSVPro) with Apnea Backup ventilation
(optional)

Ventilator (V

T

) parameter ranges

Tidal volume
range:

20 to 1500 mL (Volume Control
and SIMV modes)

Incremental
settings:

20 to 100 mL (increments of 5 mL)
100 to 300 mL (increments of 10 mL)
300 to 1000 mL (increments of 25 mL)
1000 to 1500 mL (increments of 50 mL)

Minute volume
range:

0 to 99.9 L/min

Pressure
(P

Inspired

) range: 5 to 60 cm H

2

O (increments of 1 cm H

2

O)

Pressure
(P

limit

) range:

12 to 100 cm H

2

O (increments of 1 cm H

2

O)

Pressure
(P

support

)

range:

OFF, 2 to 40 cm H

2

O (increments of 1 cm H

2

O)

Rate:

4 to 100 breaths per minute for Volume
Control and Pressure Control; 2 to 60
breaths per minute for SIMV, PSVPro and
SIMV–PC+PSV (increments of 1 breath
per minute)

Inspiratory/
expiratory ratio: 2:1 to 1:8 (increments of 0.5)

Inspiratory time: 0.2 to 5 seconds (increments of 0.1 seconds)

(SIMV and PSVPro)

Trigger window: 0 to 80% (increments of 5%)

Flow trigger:

0.2 to 1 L/min (increments of 0.2 L/min)
1 to 10 L/min (increments of 0.5 L/min)

Inspiration
termination
level:

5 to 75% (increments of 5%)

Backup
mode delay:

10 to 30 seconds (increments of 5 seconds)

Positive End Expiratory Pressure (PEEP)

Type:

Integrated, electronically controlled

Range:

OFF, 4 to 30 cm H

2

O (increments of 1 cm H

2

O)

Ventilator performance

Pressure
range at inlet:

240 kPa to 700 kPa/35 psig to 100 psig

Peak gas flow:

120 L/min + fresh gas flow

Flow valve
range:

1 to 120 L/min

Flow
compensation
range:

200 mL/min to 15 L/min

Ventilator monitoring

Expiratory
minute
volume range:

0 to 99.9 L/min

Expiratory tidal
volume range:

0 to 9999 mL

O

2

%:

8 to 100%

Peak pressure:

–20 to 120 cm H

2

O

Mean pressure: –20 to 120 cm H

2

O

Plateau
pressure:

0 to 120 cm H

2

O

Pressure
waveform
sweep speed:

2 to 25 breaths per minute (0 to 15 seconds)
26 to 75 breaths per minute (0 to 5 seconds)
75 breaths per minute (0 to 3 seconds)

AN3175-E.qxd 19.6.2006 12:37 Sivu 3