GE Healthcare Aestiva_5 MRI anesthesia machine User Manual
Page 6
Pneumatic specifications, continued
O
2
controls
Method:
Proportionate decrease of N
2
O, CO
2
, He/O
2
with reduction in O
2
pressure
Supply failure
alarm range:
193 kPa to 221 kPa/28 psig to 32 psig
Sounds at maximum volume every 10 seconds
O
2
flush range:
35 to 50 L/min
Flowmeters
O
2
ranges:
Two tubes: 0.05 to 0.95 L/min and 1 to 15 L/min
Minimum O
2
flow: 50 mL/min ±25 mL
N
2
O ranges:
Two tubes: 0 to 0.95 L/min and 1 to 10 L/min
Air range:
One tube option: 1 to 15 L/min
Two tube option: 0 to 0.95 and 1 to 15 L/min
(low flow tube optional)
CO
2
(optional):
One tube: 0 to 0.5 L/min
Heliox range
(optional):
One tube: 0 to 15 L/min
Calibration:
Percent of
Accuracy
full scale flow
(% of flowrate)
100
±2.5%
90
±2.5%
80
±2.6%
70
±2.7%
60
±2.9%
50
±3.1%
40
±3.4%
30
±4.0%
20
±5.0%
10
±8.1%
Calibration
conditions:*
20°C/68°F, 101.3 kPa/760 mmHg
*
Different breathing circuit pressures, barometric pressures or
temperatures change flowtube accuracy.
Hypoxic guard system
Type:
Mechanical Link-25
™
Range:
Provides a nominal minimum 25%
concentration of oxygen in O
2
/N
2
O mixture.
Materials
All materials in contact with patient breathing gases are
free of natural rubber latex.
Environmental specifications
System operation
Temperature:
10° to 40°C/50° to 104°F
Humidity:
15 to 95% relative humidity
(non-condensing)
Altitude:
–440 to 3565 m/500 to 800 mmHg
System storage
Temperature:
–25° to 65°C/–13° to 149°F
Humidity:
10 to 100% relative humidity
(including condensing)
Altitude:
–440 to 5860 m/375 to 800 mmHg
Oxygen cell
storage:
–15° to 50°C/5° to 122°F
10 to 95% relative humidity
500 to 800 mmHg
Safety and compliance
Immunity:
Complies with all requirements of EN 60601-1-2
Emissions:
CISPR 11 group 1 class B
Approvals:
UL 2601-1
CSA C22.2 #601.1
IEC 601-1
EN 60601-1
CE 0197