Pneumatic specifications, continued

O

2

controls

Method:

Proportionate decrease of N

2

O, CO

2

, He/O

2

with reduction in O

2

pressure

Supply failure
alarm range:

193 kPa to 221 kPa/28 psig to 32 psig
Sounds at maximum volume every 10 seconds

O

2

flush range:

35 to 50 L/min

Flowmeters

O

2

ranges:

Two tubes: 0.05 to 0.95 L/min and 1 to 15 L/min

Minimum O

2

flow: 50 mL/min ±25 mL

N

2

O ranges:

Two tubes: 0 to 0.95 L/min and 1 to 10 L/min

Air range:

One tube option: 1 to 15 L/min
Two tube option: 0 to 0.95 and 1 to 15 L/min
(low flow tube optional)

CO

2

(optional):

One tube: 0 to 0.5 L/min

Heliox range
(optional):

One tube: 0 to 15 L/min

Calibration:

Percent of

Accuracy

full scale flow

(% of flowrate)

100

±2.5%

90

±2.5%

80

±2.6%

70

±2.7%

60

±2.9%

50

±3.1%

40

±3.4%

30

±4.0%

20

±5.0%

10

±8.1%

Calibration
conditions:*

20°C/68°F, 101.3 kPa/760 mmHg

*

Different breathing circuit pressures, barometric pressures or
temperatures change flowtube accuracy.

Hypoxic guard system

Type:

Mechanical Link-25

Range:

Provides a nominal minimum 25%
concentration of oxygen in O

2

/N

2

O mixture.

Materials

All materials in contact with patient breathing gases are
free of natural rubber latex.

Environmental specifications

System operation

Temperature:

10° to 40°C/50° to 104°F

Humidity:

15 to 95% relative humidity
(non-condensing)

Altitude:

–440 to 3565 m/500 to 800 mmHg

System storage
Temperature:

–25° to 65°C/–13° to 149°F

Humidity:

10 to 100% relative humidity
(including condensing)

Altitude:

–440 to 5860 m/375 to 800 mmHg

Oxygen cell
storage:

–15° to 50°C/5° to 122°F
10 to 95% relative humidity
500 to 800 mmHg

Safety and compliance

Immunity:

Complies with all requirements of EN 60601-1-2

Emissions:

CISPR 11 group 1 class B

Approvals:

UL 2601-1
CSA C22.2 #601.1
IEC 601-1
EN 60601-1
CE 0197