beautypg.com

Environmental specifi cations, Altitude, Trends – GE Healthcare Engström Carestation User Manual

Page 8: External communications, Eview (optional), Electrical specifi cations

background image

GE Healthcare
P.O. Box 900, FIN-00031 GE, Finland
Tel. +358 10 394 11
Fax +358 9 146 3310

www.gehealthcare.com/respiratorycare

Gas supply

Single gas
operation:

Yes

Emergency
air valve:

Built-in

Oxygen supply

Pressure range: 240 to 641 kPa/35 to 94 psi

Flow:

160 L/min

Air supply

Pressure range: 240 to 641 kPa/35 to 94 psi

Flow:

160 L/min

Environmental Specifi cations

Thermal

Operating range: 10° to 40°C

Storage range:

–20° to 65°C

Humidity

Operating range: 15 to 95% RH Non-condensing

Storage range:

15 to 95% RH Non-condensing

In accordance with IEC 68-2-3

Vibration and shock

Random
vibration:

9.5 grms @ 30 min unpacked

2 to 5000 Hz

Altitude

Operating range: –440 to 3565 m/500 to 800 mmHg

Storage range:

–440 to 5860 m/375 to 800 mmHg

Trends

(continued)

Mini-Trends
parameters
are based on
the waveform
displayed:

Paw (P

peak

, P

plat

or Leak)

Flow

(MV

exp

, RR)

Volume (Spont MV or Mech MV,

Spont RR or Mech RR)

P

aux

(P

peak

)

CO

2

(EtCO

2

)

O

2

(EtO

2

, FiO

2

)

External Communications

Communication
ports:

Serial port (RS-232), RS-485 port, RS-422 port,

1 USB port, Ethernet port, Compact fl ash

card socket, nurse call

EView (optional)

Data Available: 10 snapshots

7 days of vent data

Optional breath to breath waveform data

Electrical Specifi cations

Line supply

Line voltage:

85 to 132 Vac, 47/63 Hz

190 to 264 Vac, 47/63 Hz

Power
consumption:

< 200 W

Battery supply

Back-up battery: Built-in

Type:

Lead acid gel

Battery
back-up time:

120 minutes typical, 30 minute minimum,

battery fully charged

© 2011 General Electric Company – All rights reserved.

General Electric Company reserves the right to make
changes in specifi cation and features shown herein,
or discontinue the product described at any time without
notice or obligation. Contact your GE representative for
the most current information.

GE and GE Monogram are trademarks
of General Electric Company.

Aeroneb is a registered trademark of Aerogen, Inc.

Engström, Carestation, INview and SpiroDynamics
are trademarks of Datex-Ohmeda, Inc.

GE Healthcare, a division of General Electric Company.

NOT COMMERCIALLY AVAILABLE IN ALL MARKETS.
NOT CLEARED OR APPROVED BY THE U.S. FDA.

EMEA DOC0970212 5/11
Global version DOC0927769

EngstromCarestationSpecSheet_DOC0970212_eng.indd Sec1:8

EngstromCarestationSpecSheet_DOC0970212_eng.indd Sec1:8

26.5.2011 12.04