Airway resistance compensation (arc), Mode families, Ventilator monitoring – GE Healthcare Engström Carestation User Manual
Page 6: Non-invasive ventilation (niv) (optional), Automatic patient detection (apd), 100% o, Take snapshot, Ventilator preferences
Airway Resistance Compensation (ARC)
Note: Not available in Neonatal option
Type of
compensation:
Electronic tube compensation
Compensation for: Endotracheal and tracheostomy tubes
Tube diameter:
5 to 10 mm
Level of
compensation:
25 to 100%
Mode Families
Allows user adjustment to specify certain parameters that
align with the hospital’s current ventilator usage.
Adjustable
parameters:
Flow and Inspiratory timing
Family 1:
Flow control is On/Insp. Timing is I:E
Family 2:
Flow control is Off/Insp. Timing is I:E
Family 3:
Flow control is On/Insp. Timing is T
insp
Family 4:
Flow control is Off/Insp. Timing is T
insp
Family 5:
Flow control is On/Insp. Timing is T
pause
Ventilator Monitoring
Airway pressure
–20 to +120 cm H
2
O
Patient flow
0.1 to 32 L/min
1 to 200 L/min
Tidal volume
0.5 to 1,000 mL with Neonatal
Flow
Sensor
1 to 1,000 mL without the Neonatal
Flow
Sensor
5 to 2,500 mL
Minute volume
0 to 99.9 L/min
CO
2
0 to 30%/0 to 225 mmHg
Compliance
0.1 to 150 mL/cm H
2
O
Resistance
1 to 500 cm H2O/L/s
RQ
0.6 to 1.3
VO
2
50 to 1000 mL/min
VCO
2
50 to 1000 mL/min
Rate
0 to 150 breaths per minute
(increments of 1 breath per minute)
0 to 120 breaths per minute
(increments of 1 breath per minute)
FiO
2
10 to 100%
Rapid Shallow
Breathing Index
(RSBI)
1 to 999 bpm/L
Note: Not available in Neonatal option
Non-Invasive Ventilation (NIV) (optional)
Mask ventilation: Yes
Integrated unique leak recognition algorithm
Automatic Patient Detection (APD)
Patient
re-connection: Automatic detection in standby
Detection by:
Back pressure to Bias-fl ow
100% O
2
( O
2
)
Delivers 5 to 75% above current FiO
2
setting
for ≤ 2 minutes
Delivers 100% O
2
for ≤ 2 minutes
Can be adjusted to other O
2
%
Take Snapshot
Immediate capture and storage of critical data currently on
the Engström’s display
Stored data:
3 waveform segments
Alarm messages (up to 5, currently active)
All measured parameters
All set ventilator parameters
Maximum stored
Snapshots:
10 most recent
Cursor:
Ability to cursor across waveforms
for
specifi c measured values
Ventilator Preferences
Back-up Mode: Establishes the specifi c ventilator mode
and parameters used in the event that the
ventilator switches to Back-up ventilation
ARC:
Allows control and setting of the airway
resistance compensation
Assist Control: Allows the user to turn the Assist Control
capability On or Off
Leak
Compensation: Allows the user to turn the Leak
Compensation capability On or Off
Trigger
Compensation: Allows the user to turn On or Off
compensation for fl ow triggering
TV Based
Conditions:
Allows setting between ATPD (Ambient
Temperature Pressure Dry) or BTPS
(Body Temperature Pressure Saturated)