FRC INview (optional)

Note: Not available when Neonatal patient type is selected

• Functional Residual Capacity measurement

– Wash-in and Wash-out method provides 2 separate

FRC

measurements

• FRC displayed both numerically and graphically

• The most recent 5 FRC procedures displayed

• PEEP

e

and PEEP

i

displayed with each FRC

• FRC Event Log records:

– FRC measurements

– Ventilator settings and procedures that may affect the

FRC procedure

• Programmable time intervals for automatic
FRC

measurements

PEEP INview (available with FRC INview)

Note: Not available when Neonatal patient type is selected

• Measures FRC at up to 5 different PEEP levels

• Graphic and numeric display of FRC values

• User selectable beginning and ending PEEP levels

– Ventilator evenly spaces additional PEEP levels

– PEEP levels can either increase or decrease

• PEEP

e

and PEEP

i

displayed during each FRC measurement

Lung INview (available with SpiroDynamics
and FRC INview)

Note: Not available when Neonatal patient type is selected

• Integrates SpiroDynamics and FRC INview within the
PEEP INview procedure

• Measures the amount of volume between the Dynostatic
curves at each FRC measurement

• Estimate of recruitment volume

INview Vent Calculations (optional)

Note: Not available when Neonatal patient type is selected

Data from Engström Care station and external lab results are
used to provide the following values:

• PAO

2

– Alveolar partial pressure of oxygen

• AaDO

2

–Alveolar arterial oxygen difference

• Pa/FiO

2

– Oxygenation index

• PaO

2

/PAO

2

– Alveolar arterial oxygen pressure difference

• CO – Cardiac output

• Vd/Vt – Dead space ventilation

• Vd – Dead space volume

• VA – Alveolar ventilation

Spirometry

Data source:

Ventilator or Compact Airway Module*

(M-COV, M-COVX, M-CAiOV, M-CAiOVX; E-COV,

E-COVX, E-CAiOV, E-CAiOVX)

Loop types:

Pressure-Volume, Pressure-Flow and
Flow-Volume

Saved loop:

Up to six loops can be saved

Reference loop: A saved loop can be selected as the reference

loop to compare with the current loop
being displayed

Cursor:

Freezes current loops and provides numeric
display of X and Y axis as cursor moves
across loops

Pulmonary
mechanics:

P

peak

, P

plat

, P

mean

, PEEP

e

, PEEP

i

, TV

insp

, TV

exp

, MV

insp

,

MV

exp

, Compliance and Resistance

Auxiliary Pressure

Auxiliary
pressure (P

aux

): Measured range: -20 to +120 cm H

2

O

Alarm range: 12 to 100 cm H

2

O

Purge fl ow:

Low fl ow to clear the P

aux

line, can be

turned

Off

SpiroDynamics (optional)

Note: Not available when Neonatal patient type is selected

• Tracheal Pressure – Volume loop displayed

• Dynostatic Curve displays calculated alveolar pressure

• Tracheal pressure measured via GE’s intratracheal
pressure

sensor*

– Connects directly to Engström’s auxiliary pressure port

• 3 point compliance measurement

– at 5-15% of the breath

– at 45-55% of the breath

– at 85-95% of the breath

• Store up to 6 SpiroDynamic loops

• Store up to 6 Dynostatic curves

• Overlay up to 2 separate loops and/or curves over
current

loop

• Cursor available across all displayed loops and curves

– Pressure and volume values displayed at cursor position

• P

peak

, PEEP

e

and Raw displayed

* For complete specifications, see product specification sheets.