Frc inview (optional), Peep inview (available with frc inview), Inview vent calculations (optional) – GE Healthcare Engström Carestation User Manual
Page 5: Spirometry, Auxiliary pressure, Spirodynamics (optional)
FRC INview (optional)
Note: Not available when Neonatal patient type is selected
• Functional Residual Capacity measurement
– Wash-in and Wash-out method provides 2 separate
FRC
measurements
• FRC displayed both numerically and graphically
• The most recent 5 FRC procedures displayed
• PEEP
e
and PEEP
i
displayed with each FRC
• FRC Event Log records:
– FRC measurements
– Ventilator settings and procedures that may affect the
FRC procedure
• Programmable time intervals for automatic
FRC
measurements
PEEP INview (available with FRC INview)
Note: Not available when Neonatal patient type is selected
• Measures FRC at up to 5 different PEEP levels
• Graphic and numeric display of FRC values
• User selectable beginning and ending PEEP levels
– Ventilator evenly spaces additional PEEP levels
– PEEP levels can either increase or decrease
• PEEP
e
and PEEP
i
displayed during each FRC measurement
Lung INview (available with SpiroDynamics
and FRC INview)
Note: Not available when Neonatal patient type is selected
• Integrates SpiroDynamics and FRC INview within the
PEEP INview procedure
• Measures the amount of volume between the Dynostatic
curves at each FRC measurement
• Estimate of recruitment volume
INview Vent Calculations (optional)
Note: Not available when Neonatal patient type is selected
Data from Engström Care station and external lab results are
used to provide the following values:
• PAO
2
– Alveolar partial pressure of oxygen
• AaDO
2
–Alveolar arterial oxygen difference
• Pa/FiO
2
– Oxygenation index
• PaO
2
/PAO
2
– Alveolar arterial oxygen pressure difference
• CO – Cardiac output
• Vd/Vt – Dead space ventilation
• Vd – Dead space volume
• VA – Alveolar ventilation
Spirometry
Data source:
Ventilator or Compact Airway Module*
(M-COV, M-COVX, M-CAiOV, M-CAiOVX; E-COV,
E-COVX, E-CAiOV, E-CAiOVX)
Loop types:
Pressure-Volume, Pressure-Flow and
Flow-Volume
Saved loop:
Up to six loops can be saved
Reference loop: A saved loop can be selected as the reference
loop to compare with the current loop
being displayed
Cursor:
Freezes current loops and provides numeric
display of X and Y axis as cursor moves
across loops
Pulmonary
mechanics:
P
peak
, P
plat
, P
mean
, PEEP
e
, PEEP
i
, TV
insp
, TV
exp
, MV
insp
,
MV
exp
, Compliance and Resistance
Auxiliary Pressure
Auxiliary
pressure (P
aux
): Measured range: -20 to +120 cm H
2
O
Alarm range: 12 to 100 cm H
2
O
Purge fl ow:
Low fl ow to clear the P
aux
line, can be
turned
Off
SpiroDynamics (optional)
Note: Not available when Neonatal patient type is selected
• Tracheal Pressure – Volume loop displayed
• Dynostatic Curve displays calculated alveolar pressure
• Tracheal pressure measured via GE’s intratracheal
pressure
sensor*
– Connects directly to Engström’s auxiliary pressure port
• 3 point compliance measurement
– at 5-15% of the breath
– at 45-55% of the breath
– at 85-95% of the breath
• Store up to 6 SpiroDynamic loops
• Store up to 6 Dynostatic curves
• Overlay up to 2 separate loops and/or curves over
current
loop
• Cursor available across all displayed loops and curves
– Pressure and volume values displayed at cursor position
• P
peak
, PEEP
e
and Raw displayed
* For complete specifications, see product specification sheets.