Nebulization

Nebulizer:

Aeroneb Nebulizer System built-in

Nebulizer
technology:

Electronic micro pump

Nebulizer run time: 10, 15, 20 or 30 minutes

Auto-repeat
capability:

Cycles: 1 to 10

Pause Time: 30 sec to 8 hr

1 to 5 minutes (increments of 1 minute)

5 to 55 minutes (increments of 5 minutes)

1 to 8 hours (increments of 0.5 hour)

Nebulizer volume
setting:

2.5, 3, 5 or 6 mL

Particle size:

Aeroneb Pro: average 2.1 microns MMAD

(Mean Mass Aerodynamic Diameter)

Aeroneb Solo: 3.4 microns MMAD

Residual
volume:

Aeroneb Pro: average 0.3 mL

Aeroneb Solo: average < 0.1 mL

Performance may vary depending upon the type of drug used.
For additional information contact Aerogen or drug supplier.

Pnuematic nebulizer

Flow
compensation:

1 to 4 L/min (increments of 0.5 L/min)

1 to 12 L/min (increments of 0.5 L/min)

Monitor Module

Module capacity

4 single slot or 2 double slot modules

Compact airway
module compatibility

M-C, M-CO, M-COV, M-COVX, M-CAiO,

M-CAiOV, M-CAiOVX, E-CO, E-COV,

E-COVX, E-CAiO, E-CAiOV, E-CAiOVX,

M-miniC,

E-miniC

Note: The Engström Carestation does not utilize the Ai (inhaled anesthesia)
feature of the compact airway modules at this time.

Note: The Engström Carestation does not utilize any of the
compact airway modules when the Neonatal Option is in use.

Trends

Trend data:

Set parameters and measured data

Trend styles:

Measured and graphic

Maximum trending: 14 days (336 hours)

Trend scaling:

12 min, 1h, 2h, 4h, 6h, 8h, 10h, 12h, 24h,

36h, 48h and 72h

Resolution:

1 minute intervals for most recent 12 hours,

5 minute intervals for 12 to 48 hours,

30 minute intervals after 48 hours

Mini-Trends:

Waveform values can be displayed

as a trend in a split screen view

Oxygen Monitoring

Technology:

Dynamic Paramagnetic Oxygen

monitoring

system

Life span:

Unlimited operating life due to the use of

non-depleting

technology

Screen

Display type:

30.5 cm/12 inch touch screen full color LCD
adjustable viewing angle

Waveforms
in screen:

Three at a time

Waveform
parameters:

Pressure, fl ow, volume, CO

2

, O

2

and auxiliary pressure

Graphic scaling:

Automatic scaling for optimal size or

independent

scaling

Data:

Control parameters, patient data, alarm
settings and messages

Status indicator: Ventilation mode, battery level, clock

Favorites:

23 Hyperlink shortcuts to choose from

7 selectable at one time

Monitoring Accuracy**

Pressure readings: ±2 cm H

2

0

Volume readings:

±10% or ±1 mL, whichever is greater

(with proximal neonatal fl ow sensor)

±10% or ±5 mL, whichever is greater

(nCPAP)

±10% or ±15 mL, whichever is greater

O

2

concentration

monitor:

±3%

Delivery Accuracy**

Inspired pressure
control:

±2 cm H

2

O

Oxygen – Air
mixing:

±3% V/V of setting

Tidal volume
delivery:

±10% of setting or ±1 mL,

whichever is greater

(with proximal neonatal fl ow sensor)

±10% of setting or ±5 mL,

whichever is greater

** Ventilation delivery specifications requirements:
• Operating at EN794 and ASTM F1100 patient conditions
• Operating at 21°C and at 1000 mbar ambient pressure
• All volumes are at ambient temperature and pressure, dry (ATPD)