Nebulization, Pnuematic nebulizer, Monitor module – GE Healthcare Engström Carestation User Manual
Page 7: Trends, Oxygen monitoring, Screen, Monitoring accuracy, Delivery accuracy
Nebulization
Nebulizer:
Aeroneb Nebulizer System built-in
Nebulizer
technology:
Electronic micro pump
Nebulizer run time: 10, 15, 20 or 30 minutes
Auto-repeat
capability:
Cycles: 1 to 10
Pause Time: 30 sec to 8 hr
1 to 5 minutes (increments of 1 minute)
5 to 55 minutes (increments of 5 minutes)
1 to 8 hours (increments of 0.5 hour)
Nebulizer volume
setting:
2.5, 3, 5 or 6 mL
Particle size:
Aeroneb Pro: average 2.1 microns MMAD
(Mean Mass Aerodynamic Diameter)
Aeroneb Solo: 3.4 microns MMAD
Residual
volume:
Aeroneb Pro: average 0.3 mL
Aeroneb Solo: average < 0.1 mL
Performance may vary depending upon the type of drug used.
For additional information contact Aerogen or drug supplier.
Pnuematic nebulizer
Flow
compensation:
1 to 4 L/min (increments of 0.5 L/min)
1 to 12 L/min (increments of 0.5 L/min)
Monitor Module
Module capacity
4 single slot or 2 double slot modules
Compact airway
module compatibility
M-C, M-CO, M-COV, M-COVX, M-CAiO,
M-CAiOV, M-CAiOVX, E-CO, E-COV,
E-COVX, E-CAiO, E-CAiOV, E-CAiOVX,
M-miniC,
E-miniC
Note: The Engström Carestation does not utilize the Ai (inhaled anesthesia)
feature of the compact airway modules at this time.
Note: The Engström Carestation does not utilize any of the
compact airway modules when the Neonatal Option is in use.
Trends
Trend data:
Set parameters and measured data
Trend styles:
Measured and graphic
Maximum trending: 14 days (336 hours)
Trend scaling:
12 min, 1h, 2h, 4h, 6h, 8h, 10h, 12h, 24h,
36h, 48h and 72h
Resolution:
1 minute intervals for most recent 12 hours,
5 minute intervals for 12 to 48 hours,
30 minute intervals after 48 hours
Mini-Trends:
Waveform values can be displayed
as a trend in a split screen view
Oxygen Monitoring
Technology:
Dynamic Paramagnetic Oxygen
monitoring
system
Life span:
Unlimited operating life due to the use of
non-depleting
technology
Screen
Display type:
30.5 cm/12 inch touch screen full color LCD
adjustable viewing angle
Waveforms
in screen:
Three at a time
Waveform
parameters:
Pressure, fl ow, volume, CO
2
, O
2
and auxiliary pressure
Graphic scaling:
Automatic scaling for optimal size or
independent
scaling
Data:
Control parameters, patient data, alarm
settings and messages
Status indicator: Ventilation mode, battery level, clock
Favorites:
23 Hyperlink shortcuts to choose from
7 selectable at one time
Monitoring Accuracy**
Pressure readings: ±2 cm H
2
0
Volume readings:
±10% or ±1 mL, whichever is greater
(with proximal neonatal fl ow sensor)
±10% or ±5 mL, whichever is greater
(nCPAP)
±10% or ±15 mL, whichever is greater
O
2
concentration
monitor:
±3%
Delivery Accuracy**
Inspired pressure
control:
±2 cm H
2
O
Oxygen – Air
mixing:
±3% V/V of setting
Tidal volume
delivery:
±10% of setting or ±1 mL,
whichever is greater
(with proximal neonatal fl ow sensor)
±10% of setting or ±5 mL,
whichever is greater
** Ventilation delivery specifications requirements:
• Operating at EN794 and ASTM F1100 patient conditions
• Operating at 21°C and at 1000 mbar ambient pressure
• All volumes are at ambient temperature and pressure, dry (ATPD)
EngstromCarestationSpecSheet_DOC0970212_eng.indd Sec1:7
EngstromCarestationSpecSheet_DOC0970212_eng.indd Sec1:7
26.5.2011 12.04
26.5.2011 12.04