3 error codes, Error codes, Remote interface reference – American Magnetics 135 & 136 Liquid Helium Level Instruments (CE-Marked) User Manual
Page 49

Rev. 3
37
Remote Interface Reference
Error Codes
4.3 Error Codes
The Model 135/136 returns specific error codes for invalid commands and/
or arguments. If an error condition is returned, the command is not
processed and the configuration of the instrument is not modified. The
table below provides a list of error codes, their meaning, and any
associated limits. The error codes are common to both the serial and IEEE-
488 communication options.
Error Code
Meaning
Valid Range
−
1
LO setpoint out of range
0
≤
LO
≤
LENGTH
−
2
B setpoint out of range
0
≤
B
<
A
−
3
A setpoint out of range
B < A
≤
LENGTH
−
4
HI setpoint out of range
0
≤
HI
≤
LENGTH
−
5
Attempted to set or query for
LENGTH in PERCENT units mode
−
6
Invalid argument,
value out of
maximum calibration range
1 cm
≤
value
≤
203.2 cm
−
7
INTERVAL setting out of range
0
≤
INTERVAL
≤
600 min
−
8
Unrecognized command
−
9
Invalid argument, value was negative
or non-numeric
0RGHO
0RGHO