Installation – American Magnetics 05300PS-430-601 High Stability Integrated Power Supply System User Manual
Page 40
22
Rev.
5
Installation
Power-Up Procedure
properly make a few connections between the various system
components which were disconnected to facilitate packing and
shipping.
2. Temporarily place a short across the magnet current terminals.
Often this is most easily accomplished by unfastening the heavy
cables from the magnet current leads and fastening them together.
1
This will allow rudimentary power supply checks without
energizing the superconducting magnet.
3. Energize the Model 430 Programmer by placing the power switch
in the
I (ON) position.
Note
Ensure the Model 601 energy absorber power adapters are properly
connected to the energy absorbers and their respective AC power
receptacles.
4. When prompted by the Model 430 Programmer, energize the power
supplies and press ENTER on the Model 430 Programmer.
Note
Remember to adjust the programmer Voltage Limit settings as
necessary to account for the additional voltage (5 V) required to
operate the system with a Model 601 installed.
Warning
All power supply parameters, both hardware and software, have
been set at the AMI factory. Power supply control, with the exception
of powering
ON
and
OFF
, is done by way of the AMI Model 430
Power Supply Programmer. No field adjustments or reconfiguration
of the power supply should be attempted in the field unless
specifically described in this document or recommended in writing
by an AMI Technical Support Representative.
5. Enter a stability setting of 100%. Refer to sections 3.3 on page 29
6. Verify the various setup menu values for the system (with the
exception of the stability setting, which is to be temporarily left at
100%). If the power supply system was purchased with an AMI
magnet, AMI has preset the setup menu values for proper
operation. See sections 3.3, 3.5, 3.9 and 3.10 for more discussion of
1. If the system shipped with CamLoc quick-disconnect connectors, they may be
quickly disconnected from the magnet leads and connected together.