Table 4-4, Return values and meanings for, Supply:m – American Magnetics 05300PS-430-601 High Stability Integrated Power Supply System User Manual
Page 134: Remote interface reference
116
Rev. 5
Remote Interface Reference
SETUP Configuration Commands and Queries
•
SUPPly:CURRent:MINimum?
Returns the minimum output current capacity of the power supply in
amperes. This value can be configured only via front panel operation using
the Supply submenu and is set automatically when a preset supply type is
selected.
•
SUPPly:CURRent:MAXimum?
Returns the maximum output current capacity of the power supply in
amperes. This value can be configured only via front panel operation using
the Supply submenu and is set automatically when a preset supply type is
selected.
•
SUPPly:MODE?
Returns an integer value corresponding to the voltage output mode
according to the table below:
This value can be configured only via front panel operation using the
Supply submenu and is set automatically when a preset supply type is
selected.
•
CONFigure:STABility <
percent
>
Sets the stability setting in percent.
•
STABility?
Returns the stability setting in percent.
•
CONFigure:COILconst <
value (kG/A, T/A
>
Sets the coil constant (also referred to as the field-to-current ratio) per the
selected field units. The coil constant must be set to a non-zero, positive
value in order to command or query the Model 430 Programmer in units of
field.
Table 4-4.
Return Values and Meanings for
SUPPly:MODE?
Query
Return
Value
Meaning
0
+0.000 to +5.000
1
+0.000 to +10.000
2
-5.000 to +5.000
3
-10.000 to +10.000
4
+0.000 to -5.000
5
+0.000 to +8.000