Operation

Persistent Switch Control : Initial Heating of the Switch

See section 3.6.1 on page 29 for details of the use of the

PERSIST. SWITCH

CONTROL

key.

3.13.1 Procedure for Initial Heating of the Switch

The Model 430 Programmer remembers the state of the persistent switch

during the time that the Programmer is de-energized. If the Model 430 is

turned on when its shut down state was such that the persistent switch

was heated and Programmer commanding zero current (the normal mode

after the magnet has been discharged), the following screen will be

displayed.

1

In order to charge the magnet, the persistent switch heater must be

energized. Perform the following steps.

1. Turn on the persistent switch heater by pressing the

PERSIST.

SWITCH CONTROL

key to heat the persistent switch heater.

2. After the persistent switch heater has been heated for the preset

heating time as set by the PSwitch Heated Time variable, the

display will show the default display and wait at zero current for a

command from the operator.

3.13.2 Procedure for Entering Persistent Mode

In order to enter the persistent mode of magnet operation, the user should

perform the following steps:

1. Use either automatic or manual ramping to achieve the desired

field or current in the magnet.

1. Refer to section 3.1 on page 21.

+0.00 A

P

Mode: Paused

+0.00 Vs

PSwitch Heater: OFF

+0.00 A

Mode: Heating Switch (4)

+0.00 Vs

PSwitch Heater: ON

+0.00 A

P

Mode: Paused

+0.00 Vs

PSwitch Heater: ON