162

Rev. 5

Appendix

Power Supply Details

Closed Loop Gain

Voltage Channel

0.6

Current Channel

12.5

INPUT CHARACTERISTICS

Voltage

Nominal

230 Vac

Single phase

Range

176 - 264 Vac

Frequency

Nominal

50 - 60 Hz

>65 Hz, leakage exceeds spec

Range

47 - 65 Hz

Current

176 Vac

9.5 A

Maximum

264 Vac

6.4 A

Power factor

0.99 minimum

Nominal output power, source

Efficiency

50%

Minimum, source

Switching frequency

50 KHz

PFC Stage

EMC (Electromagnetic Compatibility) Compli-
ance

EN61326-1 (1997)

Class A equipment

EMC immunity to:

ESD

EN61000-4-2

Electrostatic discharge

Radiated RF

EN61000-4-3

Radio Frequencies

EFT

EN61000-4-4

Electrical fast transient/burst

Surges

EN61000-4-5

Conducted RF

EN61000-4-6

EMC emissions

Conducted

EN61000-3-2

Harmonics

EN61000-3-3

Fluctuation & flicker

EN55011/CISPR11

0.15 to 30 MHz

Radiated

30 to 1000 MHz

Leakage current

3.5 mA

230 Vac, 47-63 Hz

Insulation coordination

Input

Installation Category II

Overvoltage Category II

Output

Installation Category II

Overvoltage Category II

Pollution degree

2

OUTPUT CHARACTERISTICS

Type of stabilizer

Voltage-current, 4-quadrant

Switch mode

Switching frequency

100 KHz

Output Stage

Source adjustment
range

Voltage

-6 V to +6 V

0 to 50°C

Current

-125 A to +125 A

Sink adjustment range

Voltage

-6 V to +6 V

0 to 50°C, recuperated energy is returned to
line

Current

-125 A to +125 A

Table A-11.

Model 4Q06125PS Electrical Specifications (Continued)

Specification

Rating / Description

Condition