11

1 Installation

Warning

Before energizing the instrument, the earth ground of the power receptacle
must be verified to be at earth potential and able to carry the rated current of
the power circuit. Using extension cords should be avoided, however, if one
must be used, ensure the ground conductor is intact and capable of carrying
the rated current.

In the event that the ground path of the instrument becomes less than sufficient
to carry the rated current of the power circuit, the instrument should be
disconnected from power, labeled as unsafe, and removed from place of
operation.

Do not operate this instrument in the presence of flammable gases. Doing so
could result in a life-threatening explosion.

Do not modify this instrument in any way. If component replacement is
required, return the instrument to AMI facilities as described in the
Troubleshooting section of this manual.

1.

Unpack the instrument

Carefully remove the instrument, sensor, oscillators and interconnecting coaxial
cables from the shipping carton and remove all packaging material. A rack mounting
kit is supplied if the instrument was purchased with the rack mount option.

Note

If there is any shipping damage, save all packing material and contact the
shipping representative to file a damage claim. Do not return the instrument
to AMI unless prior authorization has been received.

If the chassis is a table top model, place the instrument on a flat, secure surface.

Warning

Do not remove the cabinet feet and then reinsert the original screws. Doing so
could present a severe life-threatening electrical hazard. If removal of the
cabinet feet is desired, replace the original screws with screws not to exceed
1/4" in length.