Troubleshooting, Problem possible causes remedies, 18 remrest – Activeforever RemRest 901 CPAP Machine User Manual
Page 18

18 RemRest
Problem
Possible Causes
Remedies
Unit wi not run
Unit not p ugged into a
P ug unit into a “ ive” out et
“ ive” out et
Power cord not connected
Firm y attach AC or DC cord
proper y
No air comes out whi e
Air fi ter is p ugged
C ean or rep ace the fi ter
the unit is on
Air in et is b ocked
Remove obstruction or
reposition the unit so the
in et is not b ocked.
F exib e hose not connected
Connect f exib e hose to
to device.
device.
Unit p ugged in but device
Verify the “green” ight is
not turned on
on; if not, push the on/off
button
Symptoms of Obstructive Air fi ter is p ugged
C ean or rep ace the fi ter,
S eep Apnea have
remove any possib e
reoccurred
obstructions or reposition
the device
Incorrect pressure setting
Contact your care provider
Your Obstructive S eep Apnea
Contact your physician, you
conditions have changed
may need to be retested
Dry or sore eyes
The mask or nasa interface are Readjust your mask, nasa
eaking or headgear adjustment interface or headgear. If
is causing a eak
prob em persists, contact
your dea er or care provider
Troubleshooting