Olympus AU2700 User Manual
Page 60
- 60 -
8)ONLINE ILLEGAL SAMPLE KIND No. [ aa ]
(1).Disregard the test requisition information text for the sample generating the alarm.
(2).Continue to receive the test requisition information text for the other samples within the run.
(1).When receiving the test requisition information text using online, sample kind No. of the sample generating the alarm
does not meet the specification.
(2).The display of [ ] and its meaning are as follows.
aa : Sample type
----------------------------------------------------------------
__ : Normal sample
_E : Emergency sample
_P : Stat sample
U_ : Urine normal sample
UE : Urine emergency sample
UP : Urine stat sample
X_ : Other normal sample
XE : Other emergency sample
XP : Other stat sample
----------------------------------------------------------------
9)ONLINE ILLEGAL GENDER TEXT [ a ]
(1).Disregard the test requisition information text for the sample generating the alarm.
(2).Continue to receive the test requisition information text for other samples within the run.
(1).When receiving the requisition information text using online, the sex of this text is not within the specification.
(2).The display of [ ] and its meaning are as follows.
a : Received sex
10)ONLINE ILLEGAL AGE/MONTH [ aaa bb ]
(1).Disregard the test requisition information text for the sample generating the alarm.
(2).Continue to receive the test requisition information text for other samples within the run.
(1).When receiving the requisition information text using online, age or month age (age for baby before one year old) of this
text is out of the specification.
(2).The display of [ ] and its meaning are as follows.
aaa : Received age
bb : Received month age