(3.6) Others
(1) Contents and formats of text.
Item
Digit
Contents
Remarks
1. Rack No.
(Optional)
4 or 5
'0001'-'9999' or
'00001'-'99999'
STAT sample is space
When Rack no. is no setting,
Rack No. is '00000'.
•
The digit of rack no.
and cup pos. and
these using is
changeable in
[PARAMETER]
[ONLINE]
[PROTOCOL]
menu.
•
In case ,receive
T.R.I.
in batch.
2. Cup position
(Optional)
2
rack : '01'-'10'
STAT : '01'-'22'
•
This is automatically
assigned when rack
No. is in use.
3. Sample type
1
space : serum
U : urine
X : other
4. Sample No.
4
'0001'-'9999' : Routine sample
'E001'-'E999' : Emergency sample
'P001'-'P999' : Stat sample
'R001'-'R999' : Reagent blank
'A001'-'A999' : Calibration sample
'Q001'-'Q999' : QC sample
5. Sample ID
4-26
Number or character
•
Number of digits can
be changed in the
[PARAMETER]
[FORMAT]
[FORMAT
REQUISITION]
menu.
6. Original
Sample No.
4
'0001'-'9999' : normal sample
'E001'-'E999' : emergency sample
'P001'-'P999' : stat sample
For Repeat Run Results.
7. Reserved
4
Space
8. Block
1
'0'-'9' Last text is 'E'
Identification
No.