Olympus AU2700 User Manual
Page 23
- 23 -
9.ANL unit No.
(Optional)
1
‘1’: AU5400 ISE
‘1’: AU5400 unit 1
‘2’: AU5400 unit 2
‘3’: AU5400 unit 3
‘1’: AU2700
ANL unit No. is added to the
header portion of RB, ACAL,
and QC result data. Normal
sample and rerun sample, it is
added to each item. However,
in the case of an calculate
item, a space is added. In a
normal sample and repeat run,
it can be prevented from
adding a unit number by the
setup of an on-line condition
menu. This option can be
selected in the
[PARAMETER][ONLINE]
[PROTOCOL][SPECIAL]
menu.
10.Cuvette side
(Optional)
1
AU5400
‘0’ : inner cuvette (ISE Cell 1)
‘1’ : outer cuvette (ISE Cell 2)
AU2700
‘0’ : inner cuvette(ISE)
‘1’ : outer cuvette
Cuvette side is added to the
header portion of RB, ACAL,
and QC result data. Normal
sample and rerun sample, it is
added to each item. However,
in the case of an calculate
item, a space is added. In a
normal sample and rerun, it
can be prevented from adding
a unit number by the setup of
an on-line condition menu.
This option can be selected in
the
[PARAMETER][ONLINE]
[PROTOCOL][SPECIAL]
menu.
11. Sex
(Optional)
1
M : male
F : female
SP : none sex
0 : no-set
•
These items can be
selected/canceled in
the [PARAMETER]
[FORMAT]
[FORMAT
REQUISITION]
menu.
12. Age
(Optional)
3
'000'-'150'
space : no-set *1
13. Month
(Optional)
2
'00'-'11'
space : no-set *1
14. Patient
Information
(Optional)
20
number or character
15. Online
Test No.
2
'01'-'99'
•
This number can be
changed in the
[PARAMETER]
[ONLINE] [TEST
NO.]
16. Method
1
'0':normal
'1':dilution
‘2’:condense
For Repeat Run Results.