Flags list – Olympus AU2700 User Manual
Page 44
- 44 -
A.
Appendix
(1) Flags list
Flag
Contents
(
Failed to evade the contamination since the detergent for evading the contamination is not enough.
%
The sample probe is clogged with sample.
?
Calculation unable due to abnormal photometric data
#
Sample level detection error
R
Reagent level detection error
!
Calculation unable due to abnormal data
U
Reagent absorbance value at P-START of Reagent Blank run, is smaller than the
Lower limit of the Parameter.
u
Reagent absorbance value at P-START is lower than the lower limit specified
In the Parameters in routine run.
Y
Reagent absorbance value at P-END of Reagent Blank run, is greater than the
Upper limit of the Parameter.
y
Reagent absorbance value at P-END is higher than the upper limit specified
In the Parameters in routine run.
@
Abnormally high result; absorbance of every wavelength is more than 2.5.
$
No linearity validation conducted because less than 3 data obtained in the
Kinetics.
D
Too quick reaction slope in increasing kinetics, absorbance at P-START is
Higher than MAX. OD in increasing FIXED assay, or too slow reaction slope
In decreasing kinetics (= no reaction observed)
B
Too quick reaction slope in increasing kinetics, or absorbance at P-END is
Lower than MIN. OD in increasing FIXED assay.
Linearity error in kinetics.
*
Linearity error in kinetics.
&
Prozone data error (prozone check point data error)
Z
Prozone error
F
Result higher than the dynamic range specified in the Parameters
G
Result lower than the dynamic range specified in the Parameters
x
Other QC data error
1
QC result beyond the QC value range specified in the Parameters
2
Multi-rule QC data error
3
Multi-rule QC data error
4
Multi-rule QC data error
5
Multi-rule QC data error
6
Multi-rule QC data error
7
Multi-rule QC data error
p
Result beyond the panic value specified in the Parameters
T
Abnormality found in the Inter-Item Check
P
Result higher than DECIDE RANGE designated in parameters.
N
Result lower than DECIDE RANGE designated in parameters.
H
Higher than the result value range specified in the Parameters
L
Lower than the result value range specified in the Parameters
J
Result higher than the repeat run range specified in the Parameters
K
Result lower than the repeat run range specified in the Parameters
S
Result extracted for repeat run