6)ONLINE ILLEGAL SAMPLE No. [ aa: bbbbb ]
(1).Disregard the test requisition information text for the sample generating the alarm.
(2).Continue to receive the test requisition information text for other samples within the run.
(1).When receiving the test requisition information text using online, sample No. of the sample generating the alarm
does not meet the specification.
(2).The display of [ ] and its meaning are as follows.
aa : Sample type
----------------------------------------------------------------
__ : Normal sample
_E : Emergency sample
_P : Stat sample
U_ : Urine normal sample
UE : Urine emergency sample
UP : Urine stat sample
X_ : Other Normal sample
XE : Other Emergency sample
XP : Other Stat sample
----------------------------------------------------------------
bbbb : Sample No.
----------------------------------------------------------------
0001-9999 : Normal sample
E001-E999 : Emergency
P001-P999 : Stat sample
----------------------------------------------------------------