Page 56

background image

- 56 -

3)ONLINE FORMAT ERROR
<>
(1).Disregard the test requisition information text for the sample generating the alarm.
(2).Continue to receive the test requisition information text for samples following the alarm.
<>
(1).In the Online Parameter menu, submenu Set Up, the T.R.I. receive is not designated asn "YES" for the necessary parameter.
(2).The display in the [ ] and its meaning are as follows.
aaaa : Type of error
----------------------------------------------------------------
R_ __ : normal run, a normal sample
R_ _E : normal run, an emergency sample
R_ _P : normal run, a STAT sample
R_ U_ : normal run, a urine normal sample
R_ UE : normal run, a urine emergency sample
R_ UP : normal run, a urine STAT sample
R_ X_ : normal run, a Other normal sample
R_ XE : normal run, an Other emergency sample
R_ XP : normal run, a Other STAT sample
----------------------------------------------------------------
RH __ : repeat run, a normal sample
RH _E : repeat run, an emergency sample
RH _P : repeat run, a STAT sample
RH U_ : repeat run, a urine normal sample
RH UE : repeat run, a urine emergency sample
RH UP : repeat run, a urine STAT sample
RH X_ : repeat run, a Other normal sample
RH XE : repeat run, an Other emergency sample
RH XP : repeat run, a Other STAT sample
----------------------------------------------------------------
bbbbbbbbbb : Information type that is not designated
No sample No.
No ID No.