beautypg.com

Olympus AU2700 User Manual

Page 57

background image

- 57 -

4)ONLINE ILLEGAL TEXT CODE [ aa ]

(1).Disregard the test requisition information text for the sample generating the alarm.
(2).Continue to receive the test requisition information text for samples following the alarm.

(1).When receiving the test requisition information text using online, the text classification code is out of specification.
(2).The display of [ ] and its meaning are as follows.
aa :
----------------------------------------------------------------
R_ : Normal sample
RH : Repeat run sample
----------------------------------------------------------------
5)ONLINE ILLEGAL TEXT BLOCK No. [ aa -> bb ]

(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP",
the following events occur:
1).Disregard the test requisition information text for the sample generating the alarm.
2).System will not receive text for samples following the sample generating the alarm.
3).In a case when the text requisition is received in real-time, the "STOP" is ignored and the text requisition text
will be received with real-time again during the next analysis.
(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue",
the following events occur:
1).Disregard the test requisition information text for the sample generating the alarm.
2).The test requisition information text for the next sample will be received.

(1).When receiving the requisition information text using online, the block identification No. is out of specification.
(2).The display of [ ] and its meaning are as follows.
aa : Previously received block identification No.
bb : Last received block identification No.

This manual is related to the following products: