Svm2608 specifications, Svm2608 – VXI TECHNOLOGY SVM2608 User Manual
Page 20

VXI Technology, Inc.
20
SVM2608 Introduction
SVM2608
S
PECIFICATIONS
G
ENERAL
S
PECIFICATIONS
N
UMBER OF
C
HANNELS
4
S
AMPLING
R
ATE
Range
Resolution
Accuracy
0.59 Samples/s to 100.0 kSamples/s
100 ns
1%
I
NPUT
P
OWER
+5 V dc
+12 V dc
-12 V dc
500 mA
300 mA
300 mA
VME
BUS
I
NTERFACE
Address mode
Data transfer mode
A32
D16 or D32
M
EMORY
1 MSamples per channel
V
OLTAGE
M
EASUREMENT
Range
Resolution
Accuracy
±1.0 V, ±2.0 V, ±5.0 V, ±10.0 V, ±20.0 V, ±50.0 V
1/2
15
of full scale
1%
I
NPUT
I
MPEDANCE
20 & 50 V Range
1, 2, 5 & 10 V Range
200 kΩ
> 10 MΩ
R
ESISTANCE
M
EASUREMENT
*
Range
Resolution
Accuracy
100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 100% over range*
1/6,553.6 of scale
1%
*Note: Resistance measurements can only be made one channel at a time.
All resistance measurements can be made accurately up to +199% of the set range.
I
NPUT
F
ILTER
20 kHz (-3 dB)
T
RIGGER
L
EVELS
Internal
Range
Trigger level resolution
Level
accuracy
External
Range
Trigger level resolution
Level
accuracy
trigger level determined by selected voltage range (see Voltage Measurement above)
Range/2
11
1%
-10 V to +10 V
4.88 mV
1%
D
ELAYED
T
RIGGER
Range
Resolution
0 samples to (2
32
-1) samples
1 sample
T
IMEOUT
Range
Resolution
10 µs to 227 hrs
10 µs to 100 s
W
ARM
-
UP
T
IME
10 min
MTBF
80,000 hrs (based on 20% Ground Mobile, 80% Ground Fixed environment, T ≥ 52°C)