Model 5100-02-IT Combustible Gas Sensor Module
Page: 37
Device Specific Commands Summary
Command
Number
Description
130 Key
Press
131
Set Alarm Level
132
Set Warning Level
133
Set Alarm Relay Action
134
Set Warning Relay Action
135 Reset
Alarms
136 Abort
Calibration
137
Set Calibration Gas Level
138
Apply ZERO Gas
139
Apply SPAN Gas
140 RESERVED
141
Force Gas Value
142
Reset Force Gas Value
143
Read MODBUS RTU Register
144
Set Transducer Serial Number
145
Set Calibration Mode output current
146
Set Trouble Mode output current
147 RESERVED
148 RESERVED
149
Set Gas Factor
150
Write MODBUS RTU Register
Command 130: Key Press
This command mimics the front plate key inputs.
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8 1=Enter
2=Down
4=Up
8=Menu
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns same value as passed
Command 131: Set Alarm Level
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
Alarm level, range 0 to 60. Default is 60.