6712SR Refrigerated Sampler
Section 4 Standard Programming

4-4

4

NUMBER OF BOTTLES:

1 2 4 8 12 24

Select 24.

5

BOTTLE VOLUME IS

1000 ml (300-30000)

Enter 1000.

6

SUCTION LINE LENGTH

IS 10 ft

(3-99)

Enter 7.

7

TIME PACED

FLOW PACED

Select FLOW PACED.

8

FLOW BETWEEN

SAMPLE EVENTS:

2 PULSES (1-9999)

Enter the number of pulses between sample events.

9

SEQUENTIAL

BOTTLES/SAMPLE

SAMPLES/BOTTLE

Select BOTTLES/SAMPLE.

10

RUN CONTINUOUSLY?

YES NO

For this example, select NO by pressing an arrow until the option blinks.
Then, press

. Selecting YES allows the program to run indefinitely by

repeating the sample distribution. Continuous sampling assumes that filled
bottles are replaced with empty bottles at regular service intervals.

11

1 BOTTLES PER

SAMPLE EVENT (1-24)

Enter 2.

12

NO DELAY TO START

DELAYED START

CLOCK TIME

WAIT FOR PHONE CALL

Select CLOCK TIME.

13

START FLOW COUNT AT:

00:00

Enter 6, then enter 0.
This is the time at which the sampler will start the pacing countdown.

14

SELECT DAYS:

SU MO TU WE TH FR SA

DONE

Select MO, WE, and FR.
Use the arrow keys and

to select the days. When the correct days are

blinking, select DONE and press

↵.

15

PROGRAMMING COMPLETE

RUN THIS PROGRAM

NOW?

YES NO

Select NO.

Table 4-2 Standard Program: Flow-Paced Sampling, Two Bottles Per Sample (Continued)