6712SR Refrigerated Sampler
Section 6 SDI-12 Sondes
6-2
Figure 6-1 SDI-12 Sonde Parameters
6.2 Scanning For Sondes
You must be in Extended Programming to use a sonde with the
6712. The sampler can automatically scan for the presence of
sondes if you enable that option:
• From the main menu, select OTHER FUNCTIONS.
• Select HARDWARE.
• As you press
↵ to move through the hardware screens,
you will see one that looks like the following:
Select “NO SDI-12 SONDE” to scan for sondes. If one is found,
you can set it up, and the “AUTO SDI-12 SCAN” option will be
switched to yes.
Select the “AUTO SDI-12 SCAN” option to manually turn on (or
off) automatic scanning for sondes when the sampler is turned
on.
If a sonde has already been connected and set up, the hardware
screen will instead display a screen similar to the following:
SDI-12 Sonde Data Units:
Parameter
Range
Resolution
Storage Bytes
Temperature
-40 - 100 °C
0.1 °C
2
Conductivity
0 - 100 mS/cm
0.01 mS/cm
2
Specific Conductance
0 - 100 mS/cm
0.01 mS/cm
2
Total Dissolved Solids
0 - 90 g/l
0.01 g/l
2
Salinity
0 - 70 ppt
0.1 ppt
2
Dissolved Oxygen
0 - 20 mg/l
0.1 mg/l
1
pH
0 - 14
0.1
1
ORP
-999 - 999 mv
0.1 mv
2
Level
0 - 6.5279 m
0.0001 m
2
Ammonium-Nitrogen
0 - 200 mgN/l
0.1 mgN/l
2
Ammonia-Nitrogen
0 - 200 mgN/l
0.1 mgN/l
2
Nitrate-Nitrogen
0 - 200mgN/l
0.1 mgN/l
2
Turbidity
0 - 5000 NTU
0.1 NTU
2
Chloride
0 - 1000 mg/l
0.1 mg/l
2
Chlorophyll
0-400 ug/l
0.1 ug/l
2
Other
-99999.0-99999.0 0.0001
4
NO SDI-12 SONDE
AUTO SDI-12 SCAN OFF
SDI-12 DATA:
TEMP0 pH0 SD_CO0
SAL0 05DATA0 DO0
ORP0 LEVEL0 09DATA0
↑
↓
↑
↓