Command Descriptions
Option 01 VXI Interface Module User Manual
3–31
Command
Response
trig
Open the first relay and close the second relay in the
sequence.
trig
Open the second relay and close the third relay in the
sequence.
trig
Open the third relay in the sequence. Place the trigger
subsystem in the idle state.
INITiate
:CONTinuous
N/A
N/A
The trigger subsystem is placed in the IDLE state.
ABORt
[ROUTe:]SCAN
TRIGger[:SEQuence]:SOURce
TRIGger[:SEQuence]:IMMediate
TRIGger[:SEQuence]:DELay
If a scan list has been defined by a [ROUTe:]SCAN command, this command
places the trigger subsystem in the armed state. In this state, a trigger event
enabled by the TRIGger[:SEQuence]:SOURce command opens the currently
closed channel in a scan list and closes the next channel in the list. Triggers
continue to be recognized until an ABORt command is received, which returns
the trigger subsystem to the IDLE state.
In this example, it is assumed that the Option 01 daughter board is installed on a
VX4330 Module and that the default module name “m1” is in effect.
Command
Response
conf twire,m1,(1)
Set the configure section 1 of the VX4330 to 20-to-1 twire.
scan (@m1(1!1:20!1))
Define a scan list consisting of channels 1 through 20 in
section on of the VX4330 Module.
trigger:Sequence:count
2
Set the number of times to sequence through a scan list (when
an INITiate[:IMMediate] command is received) to 2.
Command Syntax
Query Syntax
Query Response
*RST Value
Related Commands
Description
Examples