Functional check – Tektronix 070-9180-01 User Manual
Page 20
Getting Started
1–10
Option 01 VXI Interface Module User Manual
Functional Check
The Option 01 VXI Interface Module executes a self test at power-on, on
direction of a VXIbus hard or soft reset condition, or on command. The
power-on self test consists of an interface self test and an instrument self test.
The self test requested by command performs only the instrument self test. A
VXIbus hard reset occurs when another device, such as the VXIbus Resource
Manager, asserts the backplane line SYSRESET*. A VXIbus soft reset occurs
when another device, such as the Option 01 commander, sets the Reset bit in the
Option 01 Control register.
During power-on, or a hard or soft reset, the following actions take place:
1. The SYSFAIL* (VME system-failure) line is set active, indicating that the
module is executing a self test, and the Failed LED is lighted. For a soft
reset, SYSFAIL* is set. All Tektronix commanders will simultaneously set
SYSFAIL INHIBIT to prevent the resource manager from prematurely
reporting the failure of a card.
2. On completion of the interface self test, SYSFAIL* is un-asserted. If the test
fails, the SYSFAIL* line remains active. If the interface self test passed, the
SYSFAIL* line is released, and the module enters the VXIbus PASSED state
(ready for normal operation). If it failed, the module enters the VXIbus
FAILED state.
3. The instrument self test, as described in the *TST? command description is
then executed. This tests the Option 01 VXI Interface daughter board and
any relay modules under its control. If the self test fails, the module makes
an internal record of the failure(s) that occur.
The default condition of the Option 01 Module after the completion of power-on
self test is as follows:
All relays on modules controlled by the Option 01 (except VX4320
Modules) are set to the normally open position.
The VX4320 RF Multiplexer module contains eight 4-to-1 RF multiplexers.
After a power-on self test is performed by the Option 01, it closes channel 1
of each section of each VX4320 that it controls.
You can run the self test at any time during normal operation by using the *TST?
command. At the end of a self test initiated by this command, the module is
restored to its pre-test state.