Tatus, Ontrol, Able – Teledyne 300M User Manual
Page 73: 4: s, Eports
Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 7-5: Control Commands
Command Message
Meaning
"C ZERO
Do a zero check
"C COMPUTE ZERO
Calibrate Zero point
1
"C SPAN
Do a span check
"C COMPUTE SPAN
Calibrate Span point
1
"C ASEQ1
Do a auto-cal sequence
2
"C ASEQ2
Do a auto-cal sequence
2
"C ASEQ3
Do a auto-cal sequence
2
"C ABORT
Aborts auto-cal sequence
"C EXITZ
Exit zero cal only
"C EXITS
Exit span cal only
"C EXIT
Exit zero span or hold
1
Executed only if the instrument is in the proper calibration mode and concentration is within
calibration limits. This command adjusts slope and offset values.
2
Initiated only If automatic calibration sequence setup is programmed and enabled.
NOTE
The commands in Table 7-5 can only be entered via the RS-232 port
when the analyzer is in the sample mode.
When a control command is issued, the CPU will respond by issuing a status report. For example
if the host computer issues the command.
C CALZ
to do a zero check, the CPU will send the status report
C DDD:HH:MM IIII START ZERO CALIBRATION
to the RS-232 output.
7-7