Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 7-5: Control Commands

Command Message

Meaning

"C ZERO"

Do a zero check

"C COMPUTE ZERO"

Calibrate Zero point

1

"C SPAN"

Do a span check

"C COMPUTE SPAN"

Calibrate Span point

1

"C ASEQ1"

Do a auto-cal sequence

2

"C ASEQ2"

Do a auto-cal sequence

2

"C ASEQ3"

Do a auto-cal sequence

2

"C ABORT"

Aborts auto-cal sequence

"C EXITZ"

Exit zero cal only

"C EXITS"

Exit span cal only

"C EXIT"

Exit zero span or hold

1

Executed only if the instrument is in the proper calibration mode and concentration is within

calibration limits. This command adjusts slope and offset values.

2

Initiated only If automatic calibration sequence setup is programmed and enabled.

NOTE

The commands in Table 7-5 can only be entered via the RS-232 port

when the analyzer is in the sample mode.

When a control command is issued, the CPU will respond by issuing a status report. For example
if the host computer issues the command.

C CALZ

to do a zero check, the CPU will send the status report

C DDD:HH:MM IIII START ZERO CALIBRATION

to the RS-232 output.

7-7