beautypg.com

Routine maintenance, Model 300m maintenance schedule, 7 flow/pressure sensor -17 – Teledyne 300M User Manual

Page 113: 11 routine maintenance, 1 model 300m maintenance schedule

background image

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

11 ROUTINE MAINTENANCE

NOTE

The operations outlined in this chapter are to be

performed by qualified maintenance personnel only.

11.1 Model 300M Maintenance Schedule

The following are the recommended periodic maintenance items for the Teledyne API Model
300M CO Analyzer:

Table 11-1: Maintenance Schedule

Date Instrument Received _____________

Item

JAN

FEB MAR APR MAY JUN

JUL AUG

SEP

OCT NOV DEC

Recommended Action

Particulate
filter element

Replace weekly or as
needed

Pump
diaphragms

Replace every 12
months

Sample flow

Check for proper flow
(800cc/min

± 20%)

annually

Dark Cal

Perform annually as
needed, or after
adjusting R7 on Sync-
Demod board.

Pneumatic
lines

Examine and clean as
necessary

Leak Check

Leak Check after
maintenance or repair
and at least annually

Clean optical
bench

As needed

11-1