Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

5.3 M300M Internal Variables

The M300M software contains many adjustable parameters. Many of the parameters are set at time
of manufacture and do not need to be adjusted for the lifetime of the instrument. It is possible to
change these variables either through the RS-232 port or the front panel. Altering the values of
many of the variables, especially those not listed on Table 5-3, w
ill adversely affect the
performance of the instrument.
Therefore it is recommended that these variables not be adjusted
unless you have a clear understanding of the effects of the change.

To access the VARS menu press SETUP-VARS-ENTR. Use the PREV-NEXT button to select the
variable of interest and press EDIT to examine/change the value, then press ENTR to save the new
value. If no change is required, press EXIT.

Table 5-3: M300M Variables

No.

Name

Units

Default
Value

Value
Range

Description

0

DAS_HOLD_OFF

MIN

15

0 - 20

DAS hold off duration after
calibration or diagnostic

1

DYN_ZERO

OFF

OFF/ON

Enable to adjust zero
calibration through remote
contact closure

2

DYN_SPAN

OFF

OFF/ON

Enable to adjust span
calibration through remote
contact closure

3

RS232_MODE

Bit Field 8

0 - 99999

Value is SUM of following
decimal numbers:

1=enable quiet mode

2=enable computer mode

4=enable security feature

8=enable front panel RS-232
menus (API protocol)

16=enable alternate protocol

32=enables multi-drop
support

64=Enable modem

4096=enable command
prompt

4 CLOCK_ADJ

Sec. 0

±60

Real-time clock adjustment

5-7