Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 10-2: Test Function Values
Test Function
Meaning
Acceptable Values
Corrective Action for
Unacceptable Values
RANGE
The Current Full Scale
Range Setting of the
analyzer's analog outputs
Any None
required
CO MEAS
The most recent detector
reading taken in Measure
mode
2500-4800 mV
Check and adjust IR source and
Sync Demodulator as described
in Section 10.6.5 and 10.6.6
CO REF
The most recent detector
reading taken in
Reference mode
2500 - 4800 mV
Check and adjust IR source and
Sync Demodulator as described
in Section 10.6.5 and 10.6.6
MR RATIO
The ratio of the CO
MEAS value to CO REF
value
1.00 - 1.5
Check CO REF and CO MEAS
values as described above.
PRES
The absolute pressure of
the sample gas in the
absorption cell
0" - 1.0" Hg below
ambient pressure
Check for pneumatic system
problems. See Section 10.6.1.
Check for pressure transducer
problems. See Section 10.6.5
SAMPLE FLOW Sample flow rate
700 - 900 sec/min
Check for pneumatic system
problems. See Section 10.6.1.
Check for flowmeter problems.
See Section 10.6.5
SAMPLE TEMP The temperature of the
sample gas in the
absorption cell
30° - 50°C
(After warm-up)
BENCH TEMP
The temperature of the
Optical Bench
50°C
(After warm-up)
WHEEL TEMP
The temperature of the
Gas Filter Wheel
68°C
(After warm-up)
BOX TEMP
The temperature inside
the analyzer chassis
up to 10°C above
ambient
DCPS
DC Power Supply
reference - A composite
of all voltages provided
by the DC Power Supply
2500
± 200 mV
10-6