Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 5-2: I/O Signals (Continued)
#
Signal
Control
Description
19
ST_HIGH_RANGE
YES
Status Bit - Autorange High Range
Logic High = M300M in high range
Logic Low = M300M in low range
20
ST_SYSTEM_OK
YES
Status Bit - System OK
Logic High = No instrument warning present
Logic Low = 1 or more alarm present
21
ST_BENCH_ALARM
YES
Status Bit - Bench Temperature Alarm
Logic High = Bench Temp out of spec
Logic Low = Bench Temp in spec
22 ST_SOURCE_ALAR
M
YES
Status Bit - IR Source Alarm
Logic High = IR Source output too low
Logic Low = IR Source output normal
23
ST_WHEEL_ALARM
YES
Status Bit - Wheel Temperature Alarm
Logic High = Wheel Temp out of spec
Logic Low = Wheel Temp in spec
24
CO_MEASURE
NO
IR detector reading during measure phase. Typically
2500 - 4500 mV.
25
CO_REFERENCE
NO
IR detector reading during reference phase. Typically
2500 - 4500 mV.
26
SAMPLE_PRESURE
NO
Sample pressure in mV. Typical sea level value =
4300 mV for 29.9" HG-A.
27 VACUUM_PRESSUR
E
NO
Sample flow in mV
28
SAMPLE_TEMP
NO
Sample temp in mV
29 BENCH_TEMP
NO
Optical Bench temp. typically 2270 mV for 48
°C
30 WHEEL_TEMP
NO
Filter Wheel temp. typically 4770 mV for 68
°C
31
BOX_TEMP
NO
Internal analyzer temp. in mV
(table continued)
5-5