beautypg.com

Oregon Scientific BPW810N User Manual

Page 4

background image

activities may alter your reading.

Wait at least an hour before taking your blood pressure
if you have just eaten a large meal.

Do not smoke or drink alcohol before taking your blood
pressure.

Do not measure your blood pressure if you are under stress.

Wait at least 3 minutes between measurements. This
allows your blood circulation to recover.

For a meaningful comparison, try to measure under
similar conditions. For example, take daily measurements
at approximately the same time, on the same wrist, or as
directed by a physician.

To stop the measurement process at any time, press .

The unit automatically switches off 1 minute after
taking a measurement. To save the battery life, press

taking

as soon as you are fi nished to turn off the unit.

To take a measurement:
1. Choose the position from which you wish to

measure - sitting or lying down.

2. Position your body so that your wrist is parallel with your

heart, using the chart and illustrations below as a guide.

3. Relax your hand and press to turn on the unit. A voice

will prompt you to sit in the correct position. Press
again. Voice will remind you to relax and remain still.
Wrist cuff will begin inflating automatically to measure
blood pressure.

NOTE

This unit can intelligently adjust the cuff pressure

and infl ate to a higher-pressure level (>180 mmHg) when

needed.

Your systolic and diastolic pressure readings fl ash on the
LCD, followed by MAP and pulse per minute readings
every 2 seconds. At the same time, your measurement
results and the blood pressure classifi cation will also be
announced.

4. To repeat the announcement, simply press TALK /

VOLUME.

5. If you want to save the record to memory, press

or

or

and the results will be saved to User 1’s or User 2’s

memory accordingly.

The measurement is stored as the fi rst (MEM 1) entry in
the user record you selected; the last entry (MEM 30) is
dropped, and all the entries in between move up 1 digit
(e.g. 28 becomes 29, and so on).

Sitting
down
with no
armrest

Place your arm across your chest with
the wrist parallel to your heart. Hold your
elbow with the other hand.

Position your wrist on a support, cushion,
or your thigh so that it is parallel to your
heart with the palm facing up.

Lying
down

IF YOU
ARE...

S i t t i n g
d o w n
with an
armrest

Place your elbow on a table, using an object
as a support under your forearm. Your wrist
should be parallel to your heart, with the
palm facing up.

THEN...

RECALLING MEASUREMENTS STORED IN MEMORY

To view a history of User 1’s records, press

.

To view a history of User 2’s records, press

.

NOTE

The most recent record (1) is shown fi rst. Each new
measurement is assigned to the fi rst (1) record. All other
records are pushed back one digit (e.g., 2 becomes 3, and
so on), and the last record (30) is dropped from the list.

Press the corresponding user button again (

or

or

)

)

to see additional records.

Press and hold the corresponding user button to auto-
advance to additional records.

The date and time of measurements taken

are

shown

with each record.

Memory records will be kept even when the batteries
become exhausted and are replaced.

DELETING A MEASUREMENT RECORD FROM MEMORY

You have the option of deleting your latest measurement
record or entire measurement history. This is useful if
measurements have not been accurately recorded and
need to be recorded again.

Problem Symptom

No
power

Display is
dim or will
not light up

Batteries are
exhausted

Replace with
new batteries

Batteries are
inserted
incorrectly

Insert the
batteries
correctly

Low
batteries

shows

on the
display

Batteries

Replace
with new
batteries

TROUBLESHOOTING

This section includes a list of error messages and frequently
asked questions for problems you may encounter with your
blood pressure monitor. If the product is not operating as you
think it should, check here before arranging for servicing.

Dimensions:
Height x Width x Depth

3.5 x 3.4 x 1.2 inches

(90 x 86 x 30.8 mm)

Weight

6.3 oz (180g) without batteries

Cuff circumference

5.3 to 7.7 inches (13.5

to 19.5 cm)

Measuring range:
Pressure 30 to 280 mmHg
(4.0 to 37.3 kPa)
Pulse 40 to 200 pulse/min

Accuracy:
Pressure +/- 3 mmHg (+/- 0.4 kPa)
Pulse +/- 5%

Power:
Power supply 3V DC, 2 LR03 / AAA /
UM4-size 1.5V batteries
Power save Auto power off after 1
minute of non-activity

Operating environment:
Operation 50°F to 104°F (10°C to 40°C)
Storage / Transport -4°F to 140°F (-20°C to 60°C)
Humidity range 10% to 83% relative humidity

BLOOD PRESSURE LOG BOOK

To create a log of your blood pressure history, complete the
personal information section at the top, then enter the details
(date, time, and measurements) for each reading you take.
To plot your history, use an S (systolic), D (diastolic) and
M (mean arterial pressure) to mark the points where each
measurement falls on the chart, then connect the points to
view your history over time.

Name: Age: Weight: (kg/lbs)

Name: Age: Weight: (kg/lbs

Name: Age: Weight: (kg/lbs

Name: Age: Weight: (kg/lbs

© 2009 Oregon Scientifi c. All rights reserved.

P/N.:300100682-00012-10

ABOUT OREGON SCIENTIFIC

Visit our website (www.oregonscientifi c.com

www.oregonscientifi c.com) to learn more

about Oregon Scientifi c products.
If you’re in the US and would like to contact our Customer
Care department directly, please visit:
www2.oregonscientifi c.com/service/support.asp

www2.oregonscientifi c.com/service/support.asp

For international inquiries, please visit:
www2.oregonscientifi c.com/about/international.asp

www2.oregonscientifi c.com/about/international.asp.

EU-DECLARATION OF CONFORMITY

Protected by US. Patent no. 7,211,047

Check This

Remedy

Hereby, Oregon Scientific, declares that this Talking Blood
Pressure Monitor(Model: BPW810N) is in compliance with
EMC directive 2004/108/EC. A copy of the signed and
dated Declaration of Conformity is available on request
via our Oregon Scientific Customer Service.

NOTE Once the pressure reaches a certain level, it will
slowly deflate until the measurement results are displayed
on the LCD.

DELETING A SINGLE RECORD

DELETING ALL RECORDS

1. Press or to recall the latest measurement

record for User 1 or User 2.

2. Press and hold SET until the LCD shows “dEL OnE”.

2. Press and hold SET until the LCD shows “dEL OnE”.

3. Press

SET. First record will be displayed on LCD

screen.

3. Press to select option to delete all records
(“dEL ALL”).

4. Press and hold SET for 2 seconds to delete all
measurement records. After you have confirmed
deleting all records, the screen shows “dEL dOnE”.

4. Press respective user button ( / ) to scroll
through records and select the record to be deleted.
5. Press and hold SET to delete record.

6. “dEL OnE” will be displayed on LCD screen. Press
and hold SET to confirm.

“dEL dOnE” indicates successful deletion.

The no --- screen indicates successful deletion and there
are no records stored.

If you decide to abort the delete action instead:

FOR

ACTION

RETURN TO SCREEN
DISPLAY

User 1 record Press twice User 1 records

displayed

User 2 records
displayed

User 2 record Press once

1. Press or to recall the latest measurement

record for User 1 or User 2.

NOTE

If you decide to abort the delete action instead:

NOTE

FOR

ACTION

RETURN TO SCREEN
DISPLAY

User 1 record Press once User 1 records

displayed

User 2 records
displayed

User 2 record Press twice

MEM

no

TECHNICAL SPECIFICATIONS

Application:
Measuring method

Oscillometric / noninvasive

Application

For adult use only

Measurement location Wrist section of lower arm
Memory

Maximum 30 records

per user
Maximum number of users

2

are low

-